Patient-Specific Unicompartmental Knee Resurfacing Arthroplasty: Use of a Novel Interference Lock to Reduce Tibial Insert Micromotion and Backside Wear



Nick Steklov1, Nam Chao*, 2, Sudesh Srivastav 3
1 Shiley Center for Orthopedic Research and Education Center (SCORE), 11025 N. Torrey Pines Road, Suite 140, LaJolla, CA, 92037, USA
2 ConforMIS, Inc., 2 Fourth Avenue, Burlington, MA 01803, USA
3 Department of Biostatistics, Tulane University, 1440 Canal Street, New Orleans, LA 70112, USA


Article Metrics

CrossRef Citations:
2
Total Statistics:

Full-Text HTML Views: 619
Abstract HTML Views: 391
PDF Downloads: 168
Total Views/Downloads: 1178
Unique Statistics:

Full-Text HTML Views: 397
Abstract HTML Views: 291
PDF Downloads: 149
Total Views/Downloads: 837



Creative Commons License
© Steklov et al.; Licensee Bentham Open.

open-access license: This is an open access article licensed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/3.0/) which permits unrestricted, non-commercial use, distribution and reproduction in any medium, provided the work is properly cited.

* Address correspondence to this author at the ConforMIS, Inc., 2 Fourth Avenue, Burlington, MA 01803, USA; Tel: (781) 345-9001; Fax: (781) 345-0147; E-mail: nam.chao@conformis.com


Abstract

Micromotion has long been associated with wear of polyethylene tibial inserts, potentially causing failure of unicompartmental knee replacement systems. One cause of micromotion is the locking mechanism between the undersurface of the polyethylene and the tibial tray. The objective of this study was to investigate the use of new novel lock designs for reducing the micromotion associated with a patient specific tibial implant. Micromotion occurring between the tibial insert and tibial tray was measured using the DVRT method for two new lock designs and compared to the micromotion measured for the prior generation lock design. In total, 18 samples were tested, six in each of the three designs: prior generation CLEARANCE lock design, new LINE-TO-LINE design, new INTERFERENCE design.

At the lowest loading of 100N, the motion index of the new interference lock design (4.9 ±2.6 microns) was less than the motion index of both the new line-to-line lock design (11.9±4.5 microns) and the prior generation clearance lock design (154.9±84.4 microns). Likewise, at the higher loading of 400N, the motion index of the new interference lock design (58.7 ±7.2 microns) was again less than the motion index of the two alternative lock designs (new line-to-line: 135.5±57.6 microns; prior generation clearance: 371.8±151.4 microns). When compared to previously published micromotion data for TKR systems, the results demonstrate that a novel interference design locking mechanism as part of a patient-specific unicompartmental knee arthroplasty system can significantly reduce micromotion that is believed to be one of the causes of osteolysis.

Keywords: polyethylene wear, micromotion, patient-specific, knee, unicompartment..