Variability in EIT Images of Lung Ventilation as a Function of Electrode Planes and Body Positions



Jie Zhang*, 1, Robert Patterson 2
1 Division of Medical Physics, Department of Radiology, University of Kentucky, Lexington, KY 40536, USA
2 Department of Physical Medicine and Rehabilitation, University of Minnesota, Minneapolis, MN 55455, USA


Article Metrics

CrossRef Citations:
7
Total Statistics:

Full-Text HTML Views: 640
Abstract HTML Views: 374
PDF Downloads: 196
Total Views/Downloads: 1210
Unique Statistics:

Full-Text HTML Views: 424
Abstract HTML Views: 240
PDF Downloads: 174
Total Views/Downloads: 838



Creative Commons License
© Zhang and Patterson; Licensee Bentham Open.

open-access license: This is an open access article licensed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/3.0/) which permits unrestricted, non-commercial use, distribution and reproduction in any medium, provided the work is properly cited.

* Address correspondence to this author at the 800 Rose St, HX311A, Lexington, KY 40536; Tel: 859-323-2146; Fax: 859-257-4457; E-mail: jnzh222@uky.edu


Abstract

This study is aimed at investigating the variability in resistivity changes in the lung region as a function of air volume, electrode plane and body position. Six normal subjects (33.8 ± 4.7 years, range from 26 to 37 years) were studied using the Sheffield Electrical Impedance Tomography (EIT) portable system. Three transverse planes at the level of second intercostal space, the level of the xiphisternal joint, and midway between upper and lower locations were chosen for measurements. For each plane, sixteen electrodes were uniformly positioned around the thorax. Data were collected with the breath held at end expiration and after inspiring 0.5, 1.0, or 1.5 liters of air from end expiration, with the subject in both the supine and sitting position. The average resistivity change in five regions, two 8x8 pixel local regions in the right lung, entire right, entire left and total lung regions, were calculated. The results show the resistivity change averaged over electrode positions and subject positions was 7-9% per liter of air, with a slightly larger resistivity change of 10 % per liter air in the lower electrode plane. There was no significant difference (p>0.05) between supine and sitting. The two 8x8 regions show a larger inter individual variability (coefficient of variation, CV, is from 30% to 382%) compared to the entire left, entire right and total lung (CV is from 11% to 51%). The results for the global regions are more consistent. The large inter individual variability appears to be a problem for clinical applications of EIT, such as regional ventilation. The variability may be mitigated by choosing appropriate electrode plane, body position and region of interest for the analysis.

Keywords: EIT, electrical impedance topography, lung, ventilation, variability..