Stability of Cachet Phakic Intraocular Lens Position During 6-Months Follow-Up



D Schiano Lomoriello*, 1, M Lombardo1, L Gualdi2, L Iacobelli3, L Tranchina4, P Ducoli1, S Serrao1
1 Fondazione G.B. Bietti IRCCS, Rome, Italy
2 DOMA, Rome, Italy
3 Istituto Neurotraumatologico Italiano, Grottaferrata, Rome, Italy
4 Dipartimento di Biopatologia e Diagnostica per Immagini, Università degli Studi “Tor Vergata”, Rome, Italy


Article Metrics

CrossRef Citations:
0
Total Statistics:

Full-Text HTML Views: 4044
Abstract HTML Views: 1996
PDF Downloads: 765
Total Views/Downloads: 6805
Unique Statistics:

Full-Text HTML Views: 1838
Abstract HTML Views: 1158
PDF Downloads: 509
Total Views/Downloads: 3505



Creative Commons License
© Schiano Lomoriello et al.; Licensee Bentham Open.

open-access license: This is an open access article licensed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by/2.5/) which permits unrestricted, non-commercial use, distribution and reproduction in any medium, provided the work is properly cited.

* Address correspondence to this author at the Via Livenza 3, 00198 Rome, Italy; Tel: +39 06 85356727; Fax: +39 06 84242333; E-mail: do.schiano@gmail.com


Abstract

Purpose:

To evaluate the position stability of a phakic intraocular lens (Cachet pIOL; Alcon, Laboratories, Inc., Fort Worth, TX) in the anterior chamber (AC) during a 6-months follow-up.

Methods:

Thirty eyes of 16 subjects underwent a Cachet pIOL implantation for the correction of high myopia from -6.50 to -16.00 D with plano targeted refraction. The position stability of the Cachet pIOL was evaluated using an Anterior Segment-OCT (AS-OCT, Carl Zeiss AG, Oberkochen, Germany) at 1- and 6-months postoperatively. Three measurements have been taken into account: 1) the minimum distance between the anterior surface of the Cachet pIOL optic disk and the central corneal endothelium; 2) the minimum distance between the optic disk’s edges of the pIOL and the endothelium; 3) the minimum distance between the posterior surface of the pIOL optic disk and the anterior surface of the crystalline lens. The endothelial cell density (ECD) was also recorded.

Results:

No statistically significant changes of the Cachet pIOL position in the AC were found during follow-up. The average changes were ≤0.1 mm between 1- and 6-months postoperatively (P>0.05). At the end of follow-up, The mean ECD loss was 0.7 % (from 2794 ± 337 cell/mm2 to 2776 ± 388 cell/mm2; P>0.05).

Conclusion:

A high position stability of the Cachet pIOL in the anterior chamber was shown during a 6-months follow-up.

Keywords: Phakic IOL, high myopia, cachet, anterior segment oct.