Osteochondral Grafting: Effect of Graft Alignment, Material Properties, and Articular Geometry



Darryl D. D’Lima*, Peter C. Chen, Clifford W. Colwell Jr.
Shiley Center for Orthopaedic Research & Education at Scripps Clinic, Scripps Clinic, La Jolla, CA, USA

Abstract

Osteochondral grafting for cartilage lesions is an attractive surgical procedure; however, the clinical results have not always been successful. Surgical recommendations differ with respect to donor site and graft placement technique. No clear biomechanical analysis of these surgical options has been reported. We hypothesized that differences in graft placement, graft biomechanical properties, and graft topography affect cartilage stresses and strains. A finite element model of articular cartilage and meniscus in a normal knee was constructed. The model was used to analyze the magnitude and the distribution of contact stresses, von Mises stresses, and compressive strains in the intact knee, after creation of an 8-mm diameter osteochondral defect, and after osteochondral grafting of the defect. The effects of graft placement, articular surface topography, and biomechanical properties were evaluated. The osteochondral defect generated minimal changes in peak contact stress (3.6 MPa) relative to the intact condition (3.4 MPa) but significantly increased peak von Mises stress (by 110%) and peak compressive strain (by 63%). A perfectly matched graft restored stresses and strains to near intact conditions. Leaving the graft proud by 0.5 mm generated the greatest increase in local stresses (peak contact stresses = 6.7 MPa). Reducing graft stiffness and curvature of articular surface had lesser effects on local stresses. Graft alignment, graft biomechanical properties, and graft topography all affected cartilage stresses and strains. Contact stresses, von Mises stresses, and compressive strains are biomechanical markers for potential tissue damage and cell death. Leaving the graft proud tends to jeopardize the graft by increasing the stresses and strains on the graft. From a biomechanical perspective, the ideal surgical procedure is a perfectly aligned graft with reasonably matched articular cartilage surface from a lower load-bearing region of the knee.



Article Information


Identifiers and Pagination:

Year: 2009
Volume: 3
First Page: 61
Last Page: 68
Publisher Id: TOORTHJ-3-61
DOI: 10.2174/1874325000903010061

Article History:

Received Date: 18/6/2009
Revision Received Date: 30/6/2009
Acceptance Date: 4/7/2009
Electronic publication date: 6/8/2009
Collection year: 2009

Article Metrics:

CrossRef Citations:
0

Total Statistics:

Full-Text HTML Views: 1334
Abstract HTML Views: 987
PDF Downloads: 253
Total Views/Downloads: 2574

Unique Statistics:

Full-Text HTML Views: 808
Abstract HTML Views: 592
PDF Downloads: 176
Total Views/Downloads: 1576
Geographical View

© D’Lima et al.; Licensee Bentham Open.

open-access license: This is an open access article licensed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/3.0/) which permits unrestricted, non-commercial use, distribution and reproduction in any medium, provided the work is properly cited.


* Address correspondence to this author at the Shiley Center for Orthopaedic Research and Education at Scripps Clinic, 11025 N. Torrey Pines Road, Suite 140, La Jolla, CA 92037, USA; Tel: 858-332-0166; Fax: 858-332-0669; E-mail: ddlima@scripps.edu




INTRODUCTION

Chondral lesions are more prevalent than previously believed. Full-thickness lesions are found in approximately 20% of knee arthroscopies and are located most commonly in the medial femoral condyle [1Hjelle K, Solheim E, Strand T, Muri R, Brittberg M. Articular cartilage defects in 1,000 knee arthroscopies Arthroscopy 2002; 18(7): 730-4.-4Widuchowski W, Widuchowski J, Trzaska T. Articular cartilage defects: study of 25,124 knee arthroscopies Knee 2007; 14(3): 177-82.]. Even partial-thickness asymptomatic cartilage defects can progress within a 2-year period resulting in the reduction in cartilage volume [5Cicuttini F, Ding C, Wluka A, Davis S, Ebeling PR, Jones G. Association of cartilage defects with loss of knee cartilage in healthy, middle-age adults: a prospective study Arthritis Rheum 2005; 52(7): 2033-9.].

Various treatment options are recommended for full-thickness chondral and osteochondral lesions, although consistent long-term clinical results are not yet available [6Alford JW, Cole BJ. Cartilage restoration, part 2: techniques, outcomes, and future directions Am J Sports Med 2005; 33(3): 443-60.]. Surgical options can be classified into repair (microfracture and abrasion arthroplasty), regeneration (ACI), and replacement (osteochondral grafting) [7Minas T, Nehrer S. Current concepts in the treatment of articular cartilage defects Orthopedics 1997; 20(6): 525-38.]. When repair results in tissue that is typically fibrocartilaginous, replacement often produces cartilage that is hyaline in nature. However, several weeks are required for the injected cells to regenerate tissue and clinical recovery is slower than with osteochondral grafting [8Horas U, Pelinkovic D, Herr G, Aigner T, Schnettler R. Autologous chondrocyte implantation and osteochondral cylinder transplantation in cartilage repair of the knee joint: a prospective, comparative trial J Bone Joint Surg Am 2003; 85-A(2): 185-92.]. Osteochondral grafting is currently the only surgical procedure that immediately replaces the lesion with native hyaline articular cartilage [9Hangody L, Feczko P, Bartha L, Bodo G, Kish G. Mosaicplasty for the treatment of articular defects of the knee and ankle Clin Orthop Relat Res 2001; (Suppl 391)S328-36.].

Several types of surgical instrumentation are available and various surgical options are recommended. One option is to create the recipient hole at a depth 2 mm less than that of the donor graft height rather than to match recipient depth with graft height [6Alford JW, Cole BJ. Cartilage restoration, part 2: techniques, outcomes, and future directions Am J Sports Med 2005; 33(3): 443-60.]. Another option is to align the graft articular surface flush with the surrounding cartilage [6Alford JW, Cole BJ. Cartilage restoration, part 2: techniques, outcomes, and future directions Am J Sports Med 2005; 33(3): 443-60.]. On the other hand, grafts that are countersunk or recessed slightly below the surrounding articular surface tend to thicken and remodel to restore articular surface congruity [10Huang FS, Simonian PT, Norman AG, Clark JM. Effects of small incongruities in a sheep model of osteochondral autografting Am J Sports Med 2004; 32(8): 1842-8.]. Yet others have speculated that a proud graft could settle to an optimum level with weight bearing [11Pearce SG, Hurtig MB, Clarnette R, Kalra M, Cowan B, Miniaci A. An investigation of 2 techniques for optimizing joint surface congruency using multiple cylindrical osteochondral autografts Arthroscopy 2001; 17(1): 50-.]. One finite element analysis, using idealized axisymmetric geometry, indicated that graft–recipient height mismatch altered joint stresses [12Wu JZ, Herzog W, Hasler EM. Inadequate placement of osteochondral plugs may induce abnormal stress-strain distributions in articular cartilage -finite element simulations Med Eng Phys 2002; 24(2): 85-97.]. An in vitro study of 4.5-mm osteochondral defects in swine knees also reported elevated peak contact pressure with graft–recipient misalignment [13Koh JL, Wirsing K, Lautenschlager E, Zhang LO. The effect of graft height mismatch on contact pressure following osteochondral grafting: a biomechanical study Am J Sports Med 2004; 32(2): 317-20.]. However, no clear biomechanical analysis has been performed on the effect of graft placement, material properties, and curvature under clinically relevant geometry and in the presence of an intact meniscus.

The finite element method was therefore used to study the effect of differences in graft placement, graft biomechanical properties, and graft topography on cartilage stresses and strains at the site of the cartilage lesion. We chose the medial femoral condyle because of the propensity for a full-thickness chondral lesion [2Curl WW, Krome J, Gordon ES, Rushing J, Smith BP, Poehling GG. Cartilage injuries: a review of 31,516 knee arthroscopies Arthroscopy 1997; 13(4): 456-60.]. We studied an 8-mm diameter defect because it was within the range of cartilage defect sizes reported during arthroscopy, was large enough for treatment to be indicated, and was small enough to require a single osteochondral graft [1Hjelle K, Solheim E, Strand T, Muri R, Brittberg M. Articular cartilage defects in 1,000 knee arthroscopies Arthroscopy 2002; 18(7): 730-4., 14Bobic V, Morgan C, Carter T. Osteochondral autologous graft transfer Oper Tech Sports Med 2000; 8: 168-78.-16Marcacci M, Kon E, Delcogliano M, Filardo G, Busacca M, Zaffagnini S. Arthroscopic autologous osteochondral grafting for cartilage defects of the knee: prospective study results at a minimum 7-year follow-up Am J Sports Med 2007; 35(12): 2014-1.].

METHODS

Geometry

MRI scans were obtained from a young adult male knee that was scanned with three-dimensional spoiled gradient echo (slice thickness = 1.5 mm). The articular surface of the femur, tibia, and meniscus was extracted by image segmentation using a commercially available software program (MIMICS, Materialise, Leuven, Belgium). For this study, the articular cartilage of the medial compartment and medial meniscus was meshed at a constant thickness of 2 mm with hexahedral elements using Hypermesh (Altair Engineering, Irvine, CA) [17Ahmad CS, Cohen ZA, Levine WN, Ateshian GA, Mow VC. Biomechanical and topographic considerations for autologous osteochondral grafting in the knee Am J Sports Med 2001; 29(2): 201-6.]. The mesh density selected resulted in an average element size of 0.5 mm x 0.5 mm x 0.5 mm (Fig. 1).

Fig. (1)

The finite element model of femoral and tibial articular cartilage and the menisci. The geometry was reconstructed from a MRI of an adult male knee.



Material Properties

Articular cartilage was modeled as a linear elastic material (Young’s modulus = 15 MPa, poisson ratio = 0.45). Loading rates during walking are too rapid to permit significant fluid flow; therefore, articular cartilage can be treated as a linear elastic material. The meniscus was modeled as a transversely isotropic material with E1 = E2 = 20 MPa in the vertical and radial directions; E3 = 150 MPa in the circumferential directions; the in-plane poisson ratio = 0.2, the out-of-plane poisson ratio = 0.3, and shear modulus = 57.7 MPa [18Haut Donahue TL, Hull ML, Rashid MM, Jacobs CR. The sensitivity of tibiofemoral contact pressure to the size and shape of the lateral and medial menisci J Orthop Res 2004; 22(4): 807-14.].

Boundary Conditions

The bottom nodes of the articular cartilage mesh were tied to rigid surfaces representing the subchondral bone of the femur and the tibia. Initial simulations with the subchondral bone modeled with stiffness of the cortical bone (20 GPa) resulted no significant differences in the contact stresses, which has also been previously reported [19Donahue TL, Hull ML, Rashid MM, Jacobs CR. A finite element model of the human knee joint for the study of tibio-femoral contact J Biomech Eng 2002; 124(3): 273-80.]. Using rigid bodies reduced the computational time for the high-density cartilage mesh necessary to achieve convergence in graft contact outcomes. Nodes at the anterior and posterior horns of the medial meniscus were also attached to the tibial subchondral springs with linear springs (net stiffness for each horn = 2000N/mm) [19Donahue TL, Hull ML, Rashid MM, Jacobs CR. A finite element model of the human knee joint for the study of tibio-femoral contact J Biomech Eng 2002; 124(3): 273-80.]. The tibia was constrained in the mediolateral and the anteroposterior directions but was allowed to translate superiorly under a compressive load of 350N. The femur was constrained in the superoinferior direction but was allowed to translate in the mediolateral and the anteroposterior directions. This reproduced the loading conditions in a human cadaver study of osteochondral defects, the results of which were used for model validation [20Guettler JH, Demetropoulos CK, Yang KH, Jurist KA. Osteochondral defects in the human knee: influence of defect size on cartilage rim stress and load redistribution to surrounding cartilage Am J Sports Med 2004; 32(6): 1451-8.]. Global frictionless contact was modeled between the femoral and tibial cartilage, between the femoral and meniscal articular surfaces, between the tibial and meniscal articular surfaces, between the graft and tibial articular surfaces, and between the sides of the graft and recipient cartilage.

Explicit Analysis

After the femoral and tibial surfaces were brought into contact, the compressive load of 350N was applied over 1 second (Fig. 1). An explicit solver (Abaqus 6.7, Dassault Systèmes, Simulia, Providence, RI) was used to solve the quasi-static analysis.

Study Conditions

The following conditions were simulated: an intact femoral condyle with and without a meniscus, a defect 8 mm in diameter, and a defect filled with a graft 8 mm in diameter. The graft position was varied at 0.25 mm increments such that the graft was either recessed below or proud above the surface of the surrounding recipient area. The elastic modulus of the graft either matched the modulus of the recipient cartilage, was decreased by 33%, or was increased by 33% to simulate the varying modulus reported for articular cartilage at different locations in the knee [21Athanasiou KA, Rosenwasser MP, Buckwalter JA, Malinin TI, Mow VC. Interspecies comparisons of in situ intrinsic mechanical properties of distal femoral cartilage J Orthop Res 1991; 9(3): 330-40.]. The topography of the graft was also varied since the topography of surface of the femoral articular cartilage varies significantly based on location. The curvature at the site of grafting for this knee was 23 and 42 m-1 in the anteroposterior and mediolateral directions, respectively; very close to the reported mean for the medial femoral condyle [17Ahmad CS, Cohen ZA, Levine WN, Ateshian GA, Mow VC. Biomechanical and topographic considerations for autologous osteochondral grafting in the knee Am J Sports Med 2001; 29(2): 201-6.]. Based on the mean values reported for typical donor sites in the trochlea and the femoral condyle, we varied the curvature of the graft surface between 5 m-1 (minimum curvature of proximal medial trochlea) and 70m-1 (maximum curvature of lateral trochlea).

RESULTS

Peak contact stresses were 3.4 MPa for the intact condition and increased slightly to 3.6 MPa in the presence of an 8-mm osteochondral defect (Fig. 2). On the other hand, peak von Mises stresses and peak principal strain increased more dramatically by 110% and 63% in the presence of an osteochondral defect. In the absence of the meniscus, the contact stresses increased more substantially to a peak of 5 MPa.

Fig. (2)

Analysis of Intact, Defect, and Meniscectomy Conditions. Inferior view of medial femoral condyle cropped to the white rectangle. Left column: Contour maps of von Mises stress distribution for the Intact condition, Defect, and Meniscectomy conditions. Right column: Peak contact stresses, von Mises stresses, and compressive strains are charted. While peak contact stress did not increase significantly in the presence of a defect, peak von Mises stress and compressive strain were substantially elevated.



As expected, replacing the osteochondral defect with a perfectly matched graft restored peak contact stresses, von Mises stresses, and compressive strains to levels very close to those seen in the intact condition (Fig. 3). Leaving the graft proud by as little as 0.25 mm significantly increased peak contact stresses, von Mises stresses, and compressive strains in the graft but reduced these measures (except for compressive strains) in the recipient area surrounding the graft. Recessing the graft tended to have the converse effect with reduction of peak contact and von Mises stresses in the graft and an increase in the adjacent recipient contact area.

Fig. (3)

Analysis of Graft surface alignment. Left column: Contour maps of von Mises stress distribution for various alignment conditions showing significant increase with proud placement of the graft. Right column: Leaving the graft proud significantly increased peak contact stresses, von Mises stresses, and compressive strains in the graft. Recessing the graft tended to reduce of peak contact and von Mises stresses in the graft (‘+’ indicates a graft raised above the recipient surface, ‘-‘ indicates a graft recessed below the recipient surface).



The material properties of articular cartilage vary based on the location within the joint. The typical donor sites (lateral margin of the trochlear groove, posterior femoral condyles, anteromedial margin of the medial femoral condyle, anterolateral portion of the intercondylar notch) have a lower elastic modulus. We therefore varied the elastic modulus of the graft by ±33%. A near linear correlation between graft stiffness and peak graft contact and von Mises stresses was found (Fig. 4).

Fig. (4)

Results of Variation in Graft Modulus. Left column: Contour maps of von Mises stress distribution for grafts of different stiffness (E = Young’s modulus). Right column: Graft stresses linearly increased with graft stiffness.



We varied the curvature of the articular surface of the graft to represent the least curvature of a graft harvested from the proximal medial trochlear region (5m-1) and the maximum curvature of lateral trochlea (70m-1); two sites that are commonly recommended as donor sites. Changing graft curvature generated a moderate increase in peak contact stresses, von Mises stresses, and compressive strain (Fig. 5). Increasing the graft curvature had a greater effect than decreasing the graft curvature.

Fig. (5)

Results of Variation in Graft Curvature. Left column: Contour maps of von Mises stress distribution show the location of peak stresses for grafts of different curvature (C = graft curvature). Right column: Increasing the graft curvature increased stresses and strains.



DISCUSSION

We designed the study to analyze the stresses and the strains generated in the graft and the recipient area under conditions simulating normal walking. We determined the sensitivity to surgical parameters, such as graft alignment, and to recipient parameters, such as articular surface curvature and biomechanical properties of the donor region. The objective was to improve surgical technique by determining the optimum mechanical conditions for graft survival and integration.

Contact stresses generated by the finite element analysis in the intact 8-mm defect and by the meniscectomy condition were validated against published experimental reports. The peak contact stress (3.4 MPa) in the intact femoral condyle at 350N compressive load was close to the experimentally measured mean in human cadaver knees (~3.5 ± 1.5 MPa) under similar loading conditions [20Guettler JH, Demetropoulos CK, Yang KH, Jurist KA. Osteochondral defects in the human knee: influence of defect size on cartilage rim stress and load redistribution to surrounding cartilage Am J Sports Med 2004; 32(6): 1451-8.]. Peak contact stress increased slightly (to 3.6 MPa) around the rim of the 8-mm diameter defect. Experimentally, no significant change in peak contact stress was reported for defects of this size [20Guettler JH, Demetropoulos CK, Yang KH, Jurist KA. Osteochondral defects in the human knee: influence of defect size on cartilage rim stress and load redistribution to surrounding cartilage Am J Sports Med 2004; 32(6): 1451-8.].

There is some debate over the relationship between the defect size and the increased contact stresses. It has been theorized that contact stresses should increase around a defect. Experiments have also shown an increase in contact stresses around larger defects (16 mm in diameter) [22Kock NB, Smolders JM, van Susante JL, Buma P, van Kampen A, Verdonschot N. A cadaveric analysis of contact stress restoration after osteochondral transplantation of a cylindrical cartilage defect Knee Surg Sports Traumatol Arthrosc 2008; 16(5): 461-8.]. On the other hand, it has been shown that peak contact stresses may be unchanged in the presence of small cartilage defects [20Guettler JH, Demetropoulos CK, Yang KH, Jurist KA. Osteochondral defects in the human knee: influence of defect size on cartilage rim stress and load redistribution to surrounding cartilage Am J Sports Med 2004; 32(6): 1451-8., 23Pena E, Calvo B, Martinez MA, Doblare M. Effect of the size and location of osteochondral defects in degenerative arthritis: a finite element simulation Comput Biol Med 2007; 37(3): 376-87.]. In the intact condition, the curved femoral condyle on a relatively flat tibial plateau concentrates contact near the centroid of the contact patch. In the presence of a defect, the contact area tends to increase and is distributed centripetally around the margins of the defect thus reducing contact stresses. Additionally, in a defect a greater proportion of the tibiofemoral load is transmitted through the meniscus away from the central tibiofemoral contact. In our model, meniscectomy increased peak contact stress by 43%, which was well within experimentally measured peak tibiofemoral contact stresses after medial meniscectomy in human cadavers (Range, 18%-235%) [24Alhalki MM, Howell SM, Hull ML. How three methods for fixing a medial meniscal autograft affect tibial contact mechanics Am J Sports Med 1999; 27(3): 320-8.-26Lee SJ, Aadalen KJ, Malaviya P, et al. Tibiofemoral contact mechanics after serial medial meniscectomies in the human cadaveric knee Am J Sports Med 2006; 34(8): 1334-44.].

Contact stresses are an important factor for surface wear but may not directly relate to subsurface stresses. Contact stresses have received more attention in part because measuring contract stresses is experimentally more feasible than measuring subsurface stresses. While contact stresses have been used as a marker for wear in artificial materials and may be somewhat related to cartilage damage, the subsurface stresses are likely to be more important in determining potential for cell death in the deeper layers, delamination and fragmentation of cartilage, and debonding from subchondral bone, which is often observed during arthroscopy. In direct contrast to the slight increase in contact stresses, the osteochondral defect generated a substantially greater increase in peak von Mises stresses (a more common measure of material yielding [27Fung YC, Tong P. Classical and computational solid mechanics In: Singapore: World Scientific 2001.]) and peak compressive strain, which has been shown to be related to chondrocyte injury [28D'Lima DD, Hashimoto S, Chen PC, Colwell CW Jr, Lotz MK. Human chondrocyte apoptosis in response to mechanical injury Osteoarthr Cartil 2001; 9(8): 712-9.]. Further, the peak compressive strains were located at the margins of the defect at the junction with subchondral bone, indicating a potential site for initiation of progressive cartilage damage. These results provide biomechanical support for the clinical report of progression of asymptomatic cartilage defects within a 2-year period [5Cicuttini F, Ding C, Wluka A, Davis S, Ebeling PR, Jones G. Association of cartilage defects with loss of knee cartilage in healthy, middle-age adults: a prospective study Arthritis Rheum 2005; 52(7): 2033-9.].

Grafting restored peak contact stress, von Mises stress, and compressive strain to near intact levels. This has been shown in previous studies [13Koh JL, Wirsing K, Lautenschlager E, Zhang LO. The effect of graft height mismatch on contact pressure following osteochondral grafting: a biomechanical study Am J Sports Med 2004; 32(2): 317-20., 22Kock NB, Smolders JM, van Susante JL, Buma P, van Kampen A, Verdonschot N. A cadaveric analysis of contact stress restoration after osteochondral transplantation of a cylindrical cartilage defect Knee Surg Sports Traumatol Arthrosc 2008; 16(5): 461-8.]. However, graft-surface alignment had a significant effect on these measures consistent with the increase in contact pressure reported in an experimental study on swine knees [13Koh JL, Wirsing K, Lautenschlager E, Zhang LO. The effect of graft height mismatch on contact pressure following osteochondral grafting: a biomechanical study Am J Sports Med 2004; 32(2): 317-20.]. When the graft was kept proud, the peak contact stress, the von Mises stress, and the compressive stain in the graft increased significantly. Although recessing the graft reduced these stresses and strains, a substantial increase was noted in the recipient area surrounding the graft. Of all the conditions tested (meniscectomy, graft alignment, graft material property, graft curvature) keeping the graft proud by 0.5 mm generated the greatest increase in stresses and strains. Even a 0.25 mm increase in graft surface over recipient surface height generated higher stresses. The peak contact stress and peak compressive strain approached levels that have been shown to induce cartilage damage and cell death [28D'Lima DD, Hashimoto S, Chen PC, Colwell CW Jr, Lotz MK. Human chondrocyte apoptosis in response to mechanical injury Osteoarthr Cartil 2001; 9(8): 712-9.-32Clements KM, Bee ZC, Crossingham GV, Adams MA, Sharif M. How severe must repetitive loading be to kill chondrocytes in articular cartilage? Osteoarthr Cartil 2001; 9(5): 499-507.]. A freshly harvested graft may be even more susceptible to damage. Note that our model does not account for graft subsidence due to collapse and compaction of the donor bone, which may occur in vivo.

Two important biomechanical factors, which can vary significantly between the donor area and the recipient site, are the material properties of the cartilage and the curvature of the articular surface [17Ahmad CS, Cohen ZA, Levine WN, Ateshian GA, Mow VC. Biomechanical and topographic considerations for autologous osteochondral grafting in the knee Am J Sports Med 2001; 29(2): 201-6., 21Athanasiou KA, Rosenwasser MP, Buckwalter JA, Malinin TI, Mow VC. Interspecies comparisons of in situ intrinsic mechanical properties of distal femoral cartilage J Orthop Res 1991; 9(3): 330-40.]. We varied the elastic modulus of the graft by ±33% relative to the recipient site and noted a general increase in contact and von Mises stresses in the graft with increasing elastic modulus. The peak compressive strain however decreased with increasing modulus, because the softer graft compressed more under otherwise identical conditions. Lower graft modulus is a clinically more relevant condition, since the donor sites are typically harvested from low load-bearing regions. Overall, reducing graft modulus did not generate detrimental stresses and strains. Altering graft curvature within the range reported for typical donor sites also produced modest changes in the stresses and strains. In general, a graft with a lower curvature, such as the proximal medial trochlea, was more forgiving than a graft with a greater curvature. This biomechanical analysis supports the use of commonly recommended donor sites despite the variations in modulus and topography.

One weakness of our model was the use of single-subject geometry. This is a common weakness of computer models, and almost all finite element models of the articular cartilage and meniscus have used a single geometry [19Donahue TL, Hull ML, Rashid MM, Jacobs CR. A finite element model of the human knee joint for the study of tibio-femoral contact J Biomech Eng 2002; 124(3): 273-80., 23Pena E, Calvo B, Martinez MA, Doblare M. Effect of the size and location of osteochondral defects in degenerative arthritis: a finite element simulation Comput Biol Med 2007; 37(3): 376-87., 33Haut Donahue TL, Hull ML, Rashid MM, Jacobs CR. How the stiffness of meniscal attachments and meniscal material properties affect tibio-femoral contact pressure computed using a validated finite element model of the human knee joint J Biomech 2003; 36(1): 19-34., 34Yao J, Snibbe J, Maloney M, Lerner AL. Stresses and strains in the medial meniscus of an ACL deficient knee under anterior loading: a finite element analysis with image-based experimental validation J Biomech Eng 2006; 128(1): 135-41.]. Condylar shape and meniscal geometry can affect stress magnitudes and distribution. However, the between-subject variability is high [20Guettler JH, Demetropoulos CK, Yang KH, Jurist KA. Osteochondral defects in the human knee: influence of defect size on cartilage rim stress and load redistribution to surrounding cartilage Am J Sports Med 2004; 32(6): 1451-8.] and requires a larger sample size to obtain statistically significant differences. While the precise magnitudes of stress and strain may vary among knees, the trends reported in our study will likely be applicable to other knee geometries and the close approximation with experimentally reported average results validates our model [20Guettler JH, Demetropoulos CK, Yang KH, Jurist KA. Osteochondral defects in the human knee: influence of defect size on cartilage rim stress and load redistribution to surrounding cartilage Am J Sports Med 2004; 32(6): 1451-8.]. Donor thickness may also affect stresses, but donor and recipient thicknesses do not vary significantly [17Ahmad CS, Cohen ZA, Levine WN, Ateshian GA, Mow VC. Biomechanical and topographic considerations for autologous osteochondral grafting in the knee Am J Sports Med 2001; 29(2): 201-6.]. Other weaknesses include using a single-defect location and using a single-sized graft. Finally, we did not study biological factors, which may be more important in graft–recipient healing.

An osteochondral defect did not result in substantially increased contact stresses, but other biomechanical markers of potential for tissue damage and cell death, such as von Mises stresses and compressive strain, were markedly increased. Leaving a graft proud tends to jeopardize the graft by increasing the potentially deleterious stresses and strains on the graft. Recessing the graft transfers the stresses to the adjacent recipient area. Reducing the elastic modulus of the graft by up to a third did not significantly affect stresses or strains. Donor–recipient mismatch in articular surface topography also modestly increased stresses and strains. From a purely biomechanical perspective, the ideal surgical procedure is a perfectly aligned graft with reasonably matched articular cartilage surface from a lower load-bearing region of the knee.

ACKNOWLEDGEMENTS

This study was funded by Scripps Health Foundation philanthropic funds.

REFERENCES

[1] Hjelle K, Solheim E, Strand T, Muri R, Brittberg M. Articular cartilage defects in 1,000 knee arthroscopies Arthroscopy 2002; 18(7): 730-4.
[2] Curl WW, Krome J, Gordon ES, Rushing J, Smith BP, Poehling GG. Cartilage injuries: a review of 31,516 knee arthroscopies Arthroscopy 1997; 13(4): 456-60.
[3] Gudas R, Kalesinskas RJ, Kimtys V, et al. A prospective randomized clinical study of mosaic osteochondral autologous transplantation versus microfracture for the treatment of osteochondral defects in the knee joint in young athletes Arthroscopy 2005; 21(9): 1066-75.
[4] Widuchowski W, Widuchowski J, Trzaska T. Articular cartilage defects: study of 25,124 knee arthroscopies Knee 2007; 14(3): 177-82.
[5] Cicuttini F, Ding C, Wluka A, Davis S, Ebeling PR, Jones G. Association of cartilage defects with loss of knee cartilage in healthy, middle-age adults: a prospective study Arthritis Rheum 2005; 52(7): 2033-9.
[6] Alford JW, Cole BJ. Cartilage restoration, part 2: techniques, outcomes, and future directions Am J Sports Med 2005; 33(3): 443-60.
[7] Minas T, Nehrer S. Current concepts in the treatment of articular cartilage defects Orthopedics 1997; 20(6): 525-38.
[8] Horas U, Pelinkovic D, Herr G, Aigner T, Schnettler R. Autologous chondrocyte implantation and osteochondral cylinder transplantation in cartilage repair of the knee joint: a prospective, comparative trial J Bone Joint Surg Am 2003; 85-A(2): 185-92.
[9] Hangody L, Feczko P, Bartha L, Bodo G, Kish G. Mosaicplasty for the treatment of articular defects of the knee and ankle Clin Orthop Relat Res 2001; (Suppl 391)S328-36.
[10] Huang FS, Simonian PT, Norman AG, Clark JM. Effects of small incongruities in a sheep model of osteochondral autografting Am J Sports Med 2004; 32(8): 1842-8.
[11] Pearce SG, Hurtig MB, Clarnette R, Kalra M, Cowan B, Miniaci A. An investigation of 2 techniques for optimizing joint surface congruency using multiple cylindrical osteochondral autografts Arthroscopy 2001; 17(1): 50-.
[12] Wu JZ, Herzog W, Hasler EM. Inadequate placement of osteochondral plugs may induce abnormal stress-strain distributions in articular cartilage -finite element simulations Med Eng Phys 2002; 24(2): 85-97.
[13] Koh JL, Wirsing K, Lautenschlager E, Zhang LO. The effect of graft height mismatch on contact pressure following osteochondral grafting: a biomechanical study Am J Sports Med 2004; 32(2): 317-20.
[14] Bobic V, Morgan C, Carter T. Osteochondral autologous graft transfer Oper Tech Sports Med 2000; 8: 168-78.
[15] Chow JC, Hantes ME, Houle JB, Zalavras CG. Arthroscopic autogenous osteochondral transplantation for treating knee cartilage defects: a 2- to 5-year follow-up study Arthroscopy 2004; 20(7): 681-90.
[16] Marcacci M, Kon E, Delcogliano M, Filardo G, Busacca M, Zaffagnini S. Arthroscopic autologous osteochondral grafting for cartilage defects of the knee: prospective study results at a minimum 7-year follow-up Am J Sports Med 2007; 35(12): 2014-1.
[17] Ahmad CS, Cohen ZA, Levine WN, Ateshian GA, Mow VC. Biomechanical and topographic considerations for autologous osteochondral grafting in the knee Am J Sports Med 2001; 29(2): 201-6.
[18] Haut Donahue TL, Hull ML, Rashid MM, Jacobs CR. The sensitivity of tibiofemoral contact pressure to the size and shape of the lateral and medial menisci J Orthop Res 2004; 22(4): 807-14.
[19] Donahue TL, Hull ML, Rashid MM, Jacobs CR. A finite element model of the human knee joint for the study of tibio-femoral contact J Biomech Eng 2002; 124(3): 273-80.
[20] Guettler JH, Demetropoulos CK, Yang KH, Jurist KA. Osteochondral defects in the human knee: influence of defect size on cartilage rim stress and load redistribution to surrounding cartilage Am J Sports Med 2004; 32(6): 1451-8.
[21] Athanasiou KA, Rosenwasser MP, Buckwalter JA, Malinin TI, Mow VC. Interspecies comparisons of in situ intrinsic mechanical properties of distal femoral cartilage J Orthop Res 1991; 9(3): 330-40.
[22] Kock NB, Smolders JM, van Susante JL, Buma P, van Kampen A, Verdonschot N. A cadaveric analysis of contact stress restoration after osteochondral transplantation of a cylindrical cartilage defect Knee Surg Sports Traumatol Arthrosc 2008; 16(5): 461-8.
[23] Pena E, Calvo B, Martinez MA, Doblare M. Effect of the size and location of osteochondral defects in degenerative arthritis: a finite element simulation Comput Biol Med 2007; 37(3): 376-87.
[24] Alhalki MM, Howell SM, Hull ML. How three methods for fixing a medial meniscal autograft affect tibial contact mechanics Am J Sports Med 1999; 27(3): 320-8.
[25] Baratz ME, Fu FH, Mengato R. Meniscal tears: the effect of meniscectomy and of repair on intraarticular contact areas and stress in the human knee: a preliminary report Am J Sports Med 1986; 14(4): 270-5.
[26] Lee SJ, Aadalen KJ, Malaviya P, et al. Tibiofemoral contact mechanics after serial medial meniscectomies in the human cadaveric knee Am J Sports Med 2006; 34(8): 1334-44.
[27] Fung YC, Tong P. Classical and computational solid mechanics In: Singapore: World Scientific 2001.
[28] D'Lima DD, Hashimoto S, Chen PC, Colwell CW Jr, Lotz MK. Human chondrocyte apoptosis in response to mechanical injury Osteoarthr Cartil 2001; 9(8): 712-9.
[29] D'Lima DD, Hashimoto S, Chen PC, Lotz MK, Colwell CW Jr. In vitro and in vivo models of cartilage injury J Bone Joint Surg Am 2001; 83-A(2 Suppl Pt 1): 22-4.
[30] D'Lima DD, Hashimoto S, Chen PC, Lotz MK, Colwell CW Jr. Cartilage injury induces chondrocyte apoptosis J Bone Joint Surg Am 2001; 83-A(2 Suppl Pt 1): 19-21.
[31] Loening AM, James IE, Levenston ME, et al. Injurious mechanical compression of bovine articular cartilage induces chondrocyte apoptosis Arch Biochem Biophys 2000; 381(2): 205-12.
[32] Clements KM, Bee ZC, Crossingham GV, Adams MA, Sharif M. How severe must repetitive loading be to kill chondrocytes in articular cartilage? Osteoarthr Cartil 2001; 9(5): 499-507.
[33] Haut Donahue TL, Hull ML, Rashid MM, Jacobs CR. How the stiffness of meniscal attachments and meniscal material properties affect tibio-femoral contact pressure computed using a validated finite element model of the human knee joint J Biomech 2003; 36(1): 19-34.
[34] Yao J, Snibbe J, Maloney M, Lerner AL. Stresses and strains in the medial meniscus of an ACL deficient knee under anterior loading: a finite element analysis with image-based experimental validation J Biomech Eng 2006; 128(1): 135-41.

Endorsements



"Open access will revolutionize 21st century knowledge work and accelerate the diffusion of ideas and evidence that support just in time learning and the evolution of thinking in a number of disciplines."


Daniel Pesut
(Indiana University School of Nursing, USA)

"It is important that students and researchers from all over the world can have easy access to relevant, high-standard and timely scientific information. This is exactly what Open Access Journals provide and this is the reason why I support this endeavor."


Jacques Descotes
(Centre Antipoison-Centre de Pharmacovigilance, France)

"Publishing research articles is the key for future scientific progress. Open Access publishing is therefore of utmost importance for wider dissemination of information, and will help serving the best interest of the scientific community."


Patrice Talaga
(UCB S.A., Belgium)

"Open access journals are a novel concept in the medical literature. They offer accessible information to a wide variety of individuals, including physicians, medical students, clinical investigators, and the general public. They are an outstanding source of medical and scientific information."


Jeffrey M. Weinberg
(St. Luke's-Roosevelt Hospital Center, USA)

"Open access journals are extremely useful for graduate students, investigators and all other interested persons to read important scientific articles and subscribe scientific journals. Indeed, the research articles span a wide range of area and of high quality. This is specially a must for researchers belonging to institutions with limited library facility and funding to subscribe scientific journals."


Debomoy K. Lahiri
(Indiana University School of Medicine, USA)

"Open access journals represent a major break-through in publishing. They provide easy access to the latest research on a wide variety of issues. Relevant and timely articles are made available in a fraction of the time taken by more conventional publishers. Articles are of uniformly high quality and written by the world's leading authorities."


Robert Looney
(Naval Postgraduate School, USA)

"Open access journals have transformed the way scientific data is published and disseminated: particularly, whilst ensuring a high quality standard and transparency in the editorial process, they have increased the access to the scientific literature by those researchers that have limited library support or that are working on small budgets."


Richard Reithinger
(Westat, USA)

"Not only do open access journals greatly improve the access to high quality information for scientists in the developing world, it also provides extra exposure for our papers."


J. Ferwerda
(University of Oxford, UK)

"Open Access 'Chemistry' Journals allow the dissemination of knowledge at your finger tips without paying for the scientific content."


Sean L. Kitson
(Almac Sciences, Northern Ireland)

"In principle, all scientific journals should have open access, as should be science itself. Open access journals are very helpful for students, researchers and the general public including people from institutions which do not have library or cannot afford to subscribe scientific journals. The articles are high standard and cover a wide area."


Hubert Wolterbeek
(Delft University of Technology, The Netherlands)

"The widest possible diffusion of information is critical for the advancement of science. In this perspective, open access journals are instrumental in fostering researches and achievements."


Alessandro Laviano
(Sapienza - University of Rome, Italy)

"Open access journals are very useful for all scientists as they can have quick information in the different fields of science."


Philippe Hernigou
(Paris University, France)

"There are many scientists who can not afford the rather expensive subscriptions to scientific journals. Open access journals offer a good alternative for free access to good quality scientific information."


Fidel Toldrá
(Instituto de Agroquimica y Tecnologia de Alimentos, Spain)

"Open access journals have become a fundamental tool for students, researchers, patients and the general public. Many people from institutions which do not have library or cannot afford to subscribe scientific journals benefit of them on a daily basis. The articles are among the best and cover most scientific areas."


M. Bendandi
(University Clinic of Navarre, Spain)

"These journals provide researchers with a platform for rapid, open access scientific communication. The articles are of high quality and broad scope."


Peter Chiba
(University of Vienna, Austria)

"Open access journals are probably one of the most important contributions to promote and diffuse science worldwide."


Jaime Sampaio
(University of Trás-os-Montes e Alto Douro, Portugal)

"Open access journals make up a new and rather revolutionary way to scientific publication. This option opens several quite interesting possibilities to disseminate openly and freely new knowledge and even to facilitate interpersonal communication among scientists."


Eduardo A. Castro
(INIFTA, Argentina)

"Open access journals are freely available online throughout the world, for you to read, download, copy, distribute, and use. The articles published in the open access journals are high quality and cover a wide range of fields."


Kenji Hashimoto
(Chiba University, Japan)

"Open Access journals offer an innovative and efficient way of publication for academics and professionals in a wide range of disciplines. The papers published are of high quality after rigorous peer review and they are Indexed in: major international databases. I read Open Access journals to keep abreast of the recent development in my field of study."


Daniel Shek
(Chinese University of Hong Kong, Hong Kong)

"It is a modern trend for publishers to establish open access journals. Researchers, faculty members, and students will be greatly benefited by the new journals of Bentham Science Publishers Ltd. in this category."


Jih Ru Hwu
(National Central University, Taiwan)


Browse Contents



Webmaster Contact: info@benthamopen.net
Copyright © 2019 Bentham Open