Mechanical Quantification of Local Bone Quality in the Humeral Head: A Feasibility Study



Alexander Scola 1, Florian Gebhard 1, Sebastian Weckbach 1, Christoph Dehner 1, Ronald Schwyn 2, Ladina Fliri 2, Götz Röderer*, 1
1 Department of Orthopaedic Trauma, Ulm University, Ulm, Germany
2 AO Research Institute, Davos, Switzerland

Abstract

Objectives:

Surgical treatment of proximal humerus fractures can be challenging due to osteoporosis. The weak bone stock makes stable implant anchorage difficult, which can result in low primary stability. Accordingly, significant failure rates, even with modern locking plates, are reported in the literature. Intraoperative knowledge of local bone quality could be helpful in improving results. This study evaluates the feasibility of local bone quality quantification using breakaway torque measurements.

Materials and Methods:

A torque measurement tool (DensiProbe™) was developed to determine local resistance to breakaway offered by the cancellous bone in the humeral head to quantify local bone quality. The tool was adapted to a standard locking plate (PHILOS, Synthes), allowing measurement in the positions of the six humeral head screws, as provided by the aiming device of the plate. Two hundred and seventy measurements were performed in 44 fresh cadaveric human humeri.

Results:

Handling of the tool was straight forward and provided reproducible results for the six different positions. The method allows discrimination between the respective positions with statistical significance, and thus provides reliable information on the local distribution of bone quality within the humeral head.

Discussion:

This study introduces a new method using breakaway torque to determine local bone quality within the humeral head in real time. Because DensiProbe is adapted to a standard locking plate, there is the potential for intraoperative application. The information provided could enable the surgeon to improve fixation of osteoporotic proximal humerus fractures.

Keywords: Bone quality, DensiProbe, fracture, osteoporosis, proximal humerus.


Article Information


Identifiers and Pagination:

Year: 2013
Volume: 7
First Page: 172
Last Page: 176
Publisher Id: TOORTHJ-7-172
DOI: 10.2174/1874325001307010172

Article History:

Received Date: 7/1/2013
Revision Received Date: 13/4/2013
Acceptance Date: 20/4/2013
Electronic publication date: 17/5/2013
Collection year: 2013

Article Metrics:

CrossRef Citations:
0

Total Statistics:

Full-Text HTML Views: 1370
Abstract HTML Views: 840
PDF Downloads: 220
Total Views/Downloads: 2430

Unique Statistics:

Full-Text HTML Views: 654
Abstract HTML Views: 454
PDF Downloads: 143
Total Views/Downloads: 1251
Geographical View

© Scola et al.; Licensee Bentham Open.

open-access license: This is an open access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.5/) which permits unrestrictive use, distribution, and reproduction in any medium, provided the original work is properly cited.


* Address correspondence to this author at the Department of Orthopaedic Trauma, Ulm University, Albert-Einstein-Allee 23, D-89081 Ulm, Germany; Tel: +49 731 50054565; Fax: +49 731 50054502; E-mail: goetz.roederer@uniklinik-ulm.de




INTRODUCTION

Treatment of proximal humerus fractures, the third most common fracture in those aged over 65 years, remains a challenge [1Baron JA, Barrett JA, Karagas MR. The epidemiology of peripheral fractures Bone 1996; 18: 209S-13S.]. The failure rate, despite the development of better implants, remains very high, being 13.7% in a study of 153 patients performed by Krappinger et al. [2Krappinger D, Bizzotto N, Riedmann S, et al. Predicting failure after surgical fixation of proximal humerus fractures Injury 2011; 42: 1283-8.]. One important reason for failure is the osteoporotic bone stock in elderly patients. The direct relationship between bone mineral density (BMD) and mechanical stability has been shown for many anatomical regions [3Lill H, Hepp P, Gowin W, et al. Age- and gender-related distribution of bone mineral density and mechanical properties of the proximal humerus Rofo 2002; 174: 1544-50.-5Tingart MJ, Lehtinen J, Zurakowski D, Warner JJ, Apreleva M. Proximal humeral fractures: regional differences in bone mineral density of the humeral head affect the fixation strength of cancellous screws J Shoulder Elbow Surg 2006; 15: 620-4.]. To assess the cancellous bone quality in the humeral head, various techniques have been used. Tingart et al. showed that the proximal half of the humeral head has a significantly higher trabecular BMD than the distal half. [5Tingart MJ, Lehtinen J, Zurakowski D, Warner JJ, Apreleva M. Proximal humeral fractures: regional differences in bone mineral density of the humeral head affect the fixation strength of cancellous screws J Shoulder Elbow Surg 2006; 15: 620-4.]. In a second investigation focusing on mechanical testing, they found that the superior anterior region has a significantly lower BMD and pull-out strength than all the other regions, and that the central region has a higher BMD compared to the inferior anterior region. Lill et al. found that the proximal aspect as well as the medial and dorsal regions of the proximal humerus have the highest BMD and strength [3Lill H, Hepp P, Gowin W, et al. Age- and gender-related distribution of bone mineral density and mechanical properties of the proximal humerus Rofo 2002; 174: 1544-50.]. High-resolution computed tomography (CT) scans of the trabecular microstructure of the proximal humerus showed that cancellous bone of the greater and lesser tuberosity presents with decreasing bone quality when analyzing increasingly deeper portions of the bone starting at the articular surface [6Kirchhoff C, Kirchhoff S, Sprecher CM, et al. X-treme CT analysis of cancellous bone at the rotator cuff insertion in human individuals with osteoporosis: superficial versus deep quality Arch Orthop Trauma Surg 2013; 133: 381-7.]. These data suggest that there may be an intraindividual variation in the distribution of local BMD within the humeral head and that standard implants may not attain the areas of best bone purchase. This could be one reason for the reported significant failure rates when using locking devices despite their superior mechanical stability having been clearly proven [7Gillespie RJ, Ramachandran V, Lea ES, Vallier HA. Biomechanical evaluation of 3-part proximal humerus fractures: a cadaveric study Orthopedics 2009; 32: 816.]. Therefore, the intraoperative knowledge of the local distribution of bone quality within the humeral head could be advantageous for the surgeon. A previous study introduced a torque measurement tool (DensiProbe™, ARI, Davos, Switzerland), which allows determination of local bone quality using mechanical torque [8Brianza S, Plecko M, Gueorguiev B, Windolf M, Schwieger K. Biomechanical evaluation of a new fixation technique for internal fixation of three-part proximal humerus fractures in a novel cadaveric model Clin Biomech (Bristol, Avon) 2010; 25: 886-92.]. A strong correlation between this method and a high resolution CT scan in the same volume of interest was shown. The purpose of the present study was to clarify whether the method provides reproducible results in a larger population and whether there is the potential for future intraoperative use.

MATERIALS AND METHODOLGY

Specimens

Forty-four unpaired fresh frozen human cadaveric humeri were processed, consisting of 21 right and 23 left humeri. Twenty specimens were of male and 24 of female origin. The mean age was 74.6 years (minimum 58 years, maximum 92 years). The bones were frozen at -20°C until processing. Prior to testing the specimens were kept overnight in a refrigerator at 4°C. The specimens were periodically sprayed with saline solution to prevent desiccation.

Instrumentation

A standard short PHILOS plate (Synthes Inc., Oberdorf, Switzerland) was mounted on the specimen in the correct surgical position as recommended by the manufacturer, using only three shaft screws. The positions of the proximal head screws, as provided by the aiming device, were numbered from 1-6 (Fig. 1). To compare the corresponding screws on the left and right sides, the positions were adapted corresponding to Table 1. Thus the placements for all the bones were adjusted to that of a right-positioned plate, with screws 1, 3, and 6 representing the posterior region and 2, 4, and 5 the anterior region of the humeral head.

Fig. (1)

PHILOS plate (Synthes Inc.) for a right proximal humerus with the numbering system of the six proximal head screws. Screws 1, 3, and 6 aim at the posterior and 2, 4, and 5 at the anterior region of the humeral head.



Table 1

Positions of the Screws on the Right Side with Corresponding Positions on the Left Side




Torques Measurements

The torque measurement tool, DensiProbe™, for the proximal humerus is a 316L medical stainless steel, 115 mm long, solid cylinder with a 2.8 mm diameter, and a 25 mm long gauging tip. The gauging tip is made of three “V-shaped” depressions, 120° apart. A solid cylinder runs through a polyacetal copolymer (POM) holder that allows rotation of the solid cylinder with negligible friction. At the opposite end of the gauging tip, a bush connection allows the attachment of a torque sensor (Fig. 2). After gently hammering the probe into the bone, the sensor is rotated while keeping the probe through the POM holder and the local resistance to breakaway offered by the cancellous bone is measured. The direct relationship to the bone quality was shown in the study of Deckelmann et al. for the spine and by Brianza et al. for the humeral head [8Brianza S, Plecko M, Gueorguiev B, Windolf M, Schwieger K. Biomechanical evaluation of a new fixation technique for internal fixation of three-part proximal humerus fractures in a novel cadaveric model Clin Biomech (Bristol, Avon) 2010; 25: 886-92., 9Deckelmann S, Schwyn R, Van der Pol B, et al. DensiProbe Spine: A Novel Instrument for Intraoperative Measurement of Bone Density in Transpedicular Screw Fixation Spine (Phila Pa 1976) 2010. [Epub ahead of print]].

Fig. (2)

DensiProbe™ (ARI, Switzerland) for determination of local bone quality using mechanical torque. In the left upper corner the tip with three “V-shaped” depressions, which is inserted in the cancellous bone, is shown. The holder (right lower corner) has a bush connection for attachment of the torque sensor.



For application via the PHILOS plate, the aiming device for the proximal head screws is mounted. The standard centering and drill sleeve are connected. A calliper is introduced in the sleeves and the distance between the flat surface on the drill sleeve and the humeral head surface is recorded. A second custom-made measuring device is used to fix stoppers on the drill at -28mm and on the torque measurement tool at -3mm to the initially recorded distance between the flat surface on the drill sleeve and the humeral head surface. This algorithm ensures that the V-shaped depression of the 25mm probe gauging tip is fully in contact with undrilled cancellous bone at a distance of 3mm (Fig. 3). A 10-Nm range torque measuring device (MECMESIN Torque Handsensor, Mecmesin, West Sussex, UK) with an accuracy of ± 0.032 Nm at 1 Nm is connected to the torque measurement tool and rotated 120° clockwise. The peak torque recorded on the sensor display is documented as an indicator of local bone resistance.

Fig. (3)

Algorithm of the torque measurement using the DensiProbe. The application is adapted to the standard tools of the PHILOS plate. A custom made caliper is used to measure the distance between the flat surface on the drill sleeve and the humeral head surface (a). The drill is set to -28mm (b, c) and the DensiProbe to -3mm (d) to that distance to ensure measurement in a 25 mm cylinder of undrilled cancellous bone at a 3mm subchondral distance (e, f).



Statistics

Statistical analysis was performed with IBM SPSS Statistics Version 21 using the Mann Whitney test based on a significance level of p ≤ 0.05. Screw positions at the same level and anterior screws versus posterior screws were evaluated for significant differences.

RESULTS

In total 270 measurements in the 44 specimens were performed. The 120° rotation always resulted in bone failure, and thus represented the maximum torque the bone could have withstood. Twenty-six measurements were excluded because of the wrong length measurement or perforation of the head. The mean torque measurement for the respective positions was 0.409 ± 0.167 Nm (position 1), 0.244 ± 0.083 Nm (position 2), 0.407 ± 0.180 Nm (position 3), 0.165 ± 0.077 Nm (position 4), 0.185 ± 0.094 Nm (position 5), and 0.379 ± 0.174 Nm (position 6) (Fig. 4, Table 2). The positions (according to Fig. 1) 4 and 5 (anteromedial or – inferior) displayed less than half the strength of positions 1, 3, and 6. Screw in position 2 had approximately half the torque of positions 1, 3, and 6. In summary, the anterior region (positions 2, 4, and 5) emphasizing anteromedial or – inferior (positions 4 and 5) of the humeral head specimen displayed the lowest trabecular stability. Statistical analysis showed highly significant differences between positions 1-2, 3-4, and 5-6 (p < 0.01). Furthermore, the analysis between anterior (positions 2, 4, and 5) and posterior (positions 1, 3, and 6) showed a highly significant difference, with p < 0.01.

Fig. (4)

Results for the torque measurements at the different screw positions with side adjustment according to Table 1. Light grey: left; dark grey: right. Positions 2, 4, and 5 represent the anterior region, having reduced trabecular stability with emphasis on the anteromedial and – inferior region (* p ≤ 0.01).



Table 2

Mean Torque Measurements at the Screw Positions with Standard Deviation




DISCUSSION

Fracture fixation of osteoporotic proximal humerus fractures remains a challenge even in the era of locking plates [10Südkamp N, Bayer J, Hepp P, et al. Open reduction and internal fixation of proximal humeral fractures with use of the locking proximal humerus plate. Results of a prospective, multicenter, observational study J Bone Joint Surg Am 2009; 91: 1320-8.,11Brunner F, Sommer C, Bahrs C, et al. Open reduction and internal fixation of proximal humerus fractures using a proximal humeral locked plate: a prospective multicenter analysis J Orthop Trauma 2009; 23: 163-72.]. Solid implant anchorage can be difficult to achieve, resulting in limited primary stability and failure of fixation [12Laflamme GY, Rouleau DM, Berry GK, et al. Percutaneous humeral plating of fractures of the proximal humerus: results of a prospective multicenter clinical trial J Orthop Trauma 2008; 22: 153-8.,13Solberg BD, Moon CN, Franco DP, Paiement GD. Surgical treatment of three and four-part proximal humeral fractures J Bone Joint Surg Am 2009; 91: 1689-97.]. One reason may be the heterogeneous distribution of bone quality within the humeral head, which can lead to screw placement in areas where the bone is not of the best quality [5Tingart MJ, Lehtinen J, Zurakowski D, Warner JJ, Apreleva M. Proximal humeral fractures: regional differences in bone mineral density of the humeral head affect the fixation strength of cancellous screws J Shoulder Elbow Surg 2006; 15: 620-4., 14Tingart MJ, Bouxsein ML, Zurakowski D, Warner JP, Apreleva M. Three-dimensional distribution of bone density in the proximal humerus Calcif Tissue Int 2003; 73: 531-6.]. However, there was no method previously available that provided information on the local bone quality within the humeral head during surgery to allow the surgeon to take this into account when performing a procedure.

The DensiProbe tool presented in this study is based on an already existing application for the femur and spine, and allows for direct measurement of local bone quality of the humeral head in real time [15Suhm N, Hengg C, Schwyn R, et al. Mechanical torque measurement predicts load to implant cut-out: a biomechanical study investigating DHS anchorage in femoral heads Arch Orthop Trauma Surg 2007; 127: 469-74.,9Deckelmann S, Schwyn R, Van der Pol B, et al. DensiProbe Spine: A Novel Instrument for Intraoperative Measurement of Bone Density in Transpedicular Screw Fixation Spine (Phila Pa 1976) 2010. [Epub ahead of print]]. The good correlation between the DensiProbe measurements and local BMD was confirmed in the study of Brianza et al. comparing the results of DensiProbe to high-resolution peripheral quantitative computed tomography (HR-pQCT) in the identical volume of interest [8Brianza S, Plecko M, Gueorguiev B, Windolf M, Schwieger K. Biomechanical evaluation of a new fixation technique for internal fixation of three-part proximal humerus fractures in a novel cadaveric model Clin Biomech (Bristol, Avon) 2010; 25: 886-92.]. This tool is able to evaluate the local bone quality at the provided screw site. The results of the present study represent a constant mapping of the local distribution of bone quality within the humeral head, and thus demonstrate that the method provides reproducible results. A comparison with existing studies regarding humeral head BMD is difficult, because of the different methods used to address the problem. Whereas the studies of Hepp et al., Lill et al., and Tingart et al. locate the BMD to anatomical regions [3Lill H, Hepp P, Gowin W, et al. Age- and gender-related distribution of bone mineral density and mechanical properties of the proximal humerus Rofo 2002; 174: 1544-50., 4Hepp P, Lill H, Bail H, et al. Where should implants be anchored in the humeral head? Clin Orthop Relat Res 2003; 415: 139-47., 5Tingart MJ, Lehtinen J, Zurakowski D, Warner JJ, Apreleva M. Proximal humeral fractures: regional differences in bone mineral density of the humeral head affect the fixation strength of cancellous screws J Shoulder Elbow Surg 2006; 15: 620-4.], our study provides a direct measurement at the local screw site. Nevertheless, the results match in certain aspects: the posterior and proximal regions of the humeral head display a better local bone property than the anterior regions [3Lill H, Hepp P, Gowin W, et al. Age- and gender-related distribution of bone mineral density and mechanical properties of the proximal humerus Rofo 2002; 174: 1544-50.-5Tingart MJ, Lehtinen J, Zurakowski D, Warner JJ, Apreleva M. Proximal humeral fractures: regional differences in bone mineral density of the humeral head affect the fixation strength of cancellous screws J Shoulder Elbow Surg 2006; 15: 620-4., 14Tingart MJ, Bouxsein ML, Zurakowski D, Warner JP, Apreleva M. Three-dimensional distribution of bone density in the proximal humerus Calcif Tissue Int 2003; 73: 531-6.]. The DensiProbe can also distinguish between different positions, such that screws aimed at the posterior region encounter a statistically significantly higher local bone quality than those aimed at the anterior region. These measurements were again consistent in all specimens. An aspect of the method which should be mentioned is the potential influence of the DensiProbe on the surrounding cancellous bone when being applied. However, its diameter is limited to its core diameter of 2.8 mm, which represents the same amount of bone destroyed as for regular pre-drilling for the screws, thus its impact, if any, would be very limited.

Although the mechanical resistance of the bone is mainly determined by the BMD, the overall bone strength depends on additional factors, including trabecular anisotropy and spacing or the remodelling state [15Suhm N, Hengg C, Schwyn R, et al. Mechanical torque measurement predicts load to implant cut-out: a biomechanical study investigating DHS anchorage in femoral heads Arch Orthop Trauma Surg 2007; 127: 469-74.-18McNamara LM, Prendergast PJ, Schaffler MB. Bone tissue material properties are altered during osteoporosis J Musculoskelet Neuronal Interact 2005; 5: 342-.]. Using a local measurement of the breakaway torque, all these aspects are taken into account and thus provide a more specific knowledge of local bone quality, or rather properties, than from BMD. With this knowledge it is possible to take into consideration the positions where bone strength is critical and thus try to improve implant design or local stability, such as using cement augmentation of the screws. Furthermore, with the intraoperative use of this tool potentially in combination with polyaxial implants, regions with better local bone resistance could be identified and results improved [19Röderer G, Erhardt J, Kuster M, et al. Second generation locked plating of proximal humerus fractures--a prospective multicentre observational study Int Orthop 2011; 35: 425-32.,20Röderer G, Erhardt J, Graf M, Kinzl L, Gebhard F. Clinical results for minimally invasive locked plating of proximal humerus fractures J Orthop Trauma 2010; 24: 400-6.].

CONCLUSION

The DensiProbe displayed consistent and reproducible results in this large sample, providing a mapping of local bone quality distribution within the humeral head, being highest in the posterior region. This method is able to distinguish between different positions within the humeral head with statistical significance. The clinical potential is considerable because it offers the possibility of real time local assessment of bone quality during surgery and could enable surgeons to adapt their procedure to the individual properties of the patient, including with cement augmentation of the screws or using polyaxial implants.

CONFLICT OF INTEREST

The authors confirm that this article content has no conflict of interest.

ACKNOWLEDGEMENTS

Declared none.

REFERENCES

[1] Baron JA, Barrett JA, Karagas MR. The epidemiology of peripheral fractures Bone 1996; 18: 209S-13S.
[2] Krappinger D, Bizzotto N, Riedmann S, et al. Predicting failure after surgical fixation of proximal humerus fractures Injury 2011; 42: 1283-8.
[3] Lill H, Hepp P, Gowin W, et al. Age- and gender-related distribution of bone mineral density and mechanical properties of the proximal humerus Rofo 2002; 174: 1544-50.
[4] Hepp P, Lill H, Bail H, et al. Where should implants be anchored in the humeral head? Clin Orthop Relat Res 2003; 415: 139-47.
[5] Tingart MJ, Lehtinen J, Zurakowski D, Warner JJ, Apreleva M. Proximal humeral fractures: regional differences in bone mineral density of the humeral head affect the fixation strength of cancellous screws J Shoulder Elbow Surg 2006; 15: 620-4.
[6] Kirchhoff C, Kirchhoff S, Sprecher CM, et al. X-treme CT analysis of cancellous bone at the rotator cuff insertion in human individuals with osteoporosis: superficial versus deep quality Arch Orthop Trauma Surg 2013; 133: 381-7.
[7] Gillespie RJ, Ramachandran V, Lea ES, Vallier HA. Biomechanical evaluation of 3-part proximal humerus fractures: a cadaveric study Orthopedics 2009; 32: 816.
[8] Brianza S, Plecko M, Gueorguiev B, Windolf M, Schwieger K. Biomechanical evaluation of a new fixation technique for internal fixation of three-part proximal humerus fractures in a novel cadaveric model Clin Biomech (Bristol, Avon) 2010; 25: 886-92.
[9] Deckelmann S, Schwyn R, Van der Pol B, et al. DensiProbe Spine: A Novel Instrument for Intraoperative Measurement of Bone Density in Transpedicular Screw Fixation Spine (Phila Pa 1976) 2010. [Epub ahead of print]
[10] Südkamp N, Bayer J, Hepp P, et al. Open reduction and internal fixation of proximal humeral fractures with use of the locking proximal humerus plate. Results of a prospective, multicenter, observational study J Bone Joint Surg Am 2009; 91: 1320-8.
[11] Brunner F, Sommer C, Bahrs C, et al. Open reduction and internal fixation of proximal humerus fractures using a proximal humeral locked plate: a prospective multicenter analysis J Orthop Trauma 2009; 23: 163-72.
[12] Laflamme GY, Rouleau DM, Berry GK, et al. Percutaneous humeral plating of fractures of the proximal humerus: results of a prospective multicenter clinical trial J Orthop Trauma 2008; 22: 153-8.
[13] Solberg BD, Moon CN, Franco DP, Paiement GD. Surgical treatment of three and four-part proximal humeral fractures J Bone Joint Surg Am 2009; 91: 1689-97.
[14] Tingart MJ, Bouxsein ML, Zurakowski D, Warner JP, Apreleva M. Three-dimensional distribution of bone density in the proximal humerus Calcif Tissue Int 2003; 73: 531-6.
[15] Suhm N, Hengg C, Schwyn R, et al. Mechanical torque measurement predicts load to implant cut-out: a biomechanical study investigating DHS anchorage in femoral heads Arch Orthop Trauma Surg 2007; 127: 469-74.
[16] Ulrich D, van Rietbergen B, Laib A, Ruegsegger P. The ability of three-dimensional structural indices to reflect mechanical aspects of trabecular bone Bone 1999; 25: 55-60.
[17] Homminga J, McCreadie BR, Ciarelli TE, et al. Cancellous bone mechanical properties from normals and patients with hip fractures differ on the structure level, not on the bone hard tissue level Bone 2002; 30: 759-64.
[18] McNamara LM, Prendergast PJ, Schaffler MB. Bone tissue material properties are altered during osteoporosis J Musculoskelet Neuronal Interact 2005; 5: 342-.
[19] Röderer G, Erhardt J, Kuster M, et al. Second generation locked plating of proximal humerus fractures--a prospective multicentre observational study Int Orthop 2011; 35: 425-32.
[20] Röderer G, Erhardt J, Graf M, Kinzl L, Gebhard F. Clinical results for minimally invasive locked plating of proximal humerus fractures J Orthop Trauma 2010; 24: 400-6.

Endorsements



"Open access will revolutionize 21st century knowledge work and accelerate the diffusion of ideas and evidence that support just in time learning and the evolution of thinking in a number of disciplines."


Daniel Pesut
(Indiana University School of Nursing, USA)

"It is important that students and researchers from all over the world can have easy access to relevant, high-standard and timely scientific information. This is exactly what Open Access Journals provide and this is the reason why I support this endeavor."


Jacques Descotes
(Centre Antipoison-Centre de Pharmacovigilance, France)

"Publishing research articles is the key for future scientific progress. Open Access publishing is therefore of utmost importance for wider dissemination of information, and will help serving the best interest of the scientific community."


Patrice Talaga
(UCB S.A., Belgium)

"Open access journals are a novel concept in the medical literature. They offer accessible information to a wide variety of individuals, including physicians, medical students, clinical investigators, and the general public. They are an outstanding source of medical and scientific information."


Jeffrey M. Weinberg
(St. Luke's-Roosevelt Hospital Center, USA)

"Open access journals are extremely useful for graduate students, investigators and all other interested persons to read important scientific articles and subscribe scientific journals. Indeed, the research articles span a wide range of area and of high quality. This is specially a must for researchers belonging to institutions with limited library facility and funding to subscribe scientific journals."


Debomoy K. Lahiri
(Indiana University School of Medicine, USA)

"Open access journals represent a major break-through in publishing. They provide easy access to the latest research on a wide variety of issues. Relevant and timely articles are made available in a fraction of the time taken by more conventional publishers. Articles are of uniformly high quality and written by the world's leading authorities."


Robert Looney
(Naval Postgraduate School, USA)

"Open access journals have transformed the way scientific data is published and disseminated: particularly, whilst ensuring a high quality standard and transparency in the editorial process, they have increased the access to the scientific literature by those researchers that have limited library support or that are working on small budgets."


Richard Reithinger
(Westat, USA)

"Not only do open access journals greatly improve the access to high quality information for scientists in the developing world, it also provides extra exposure for our papers."


J. Ferwerda
(University of Oxford, UK)

"Open Access 'Chemistry' Journals allow the dissemination of knowledge at your finger tips without paying for the scientific content."


Sean L. Kitson
(Almac Sciences, Northern Ireland)

"In principle, all scientific journals should have open access, as should be science itself. Open access journals are very helpful for students, researchers and the general public including people from institutions which do not have library or cannot afford to subscribe scientific journals. The articles are high standard and cover a wide area."


Hubert Wolterbeek
(Delft University of Technology, The Netherlands)

"The widest possible diffusion of information is critical for the advancement of science. In this perspective, open access journals are instrumental in fostering researches and achievements."


Alessandro Laviano
(Sapienza - University of Rome, Italy)

"Open access journals are very useful for all scientists as they can have quick information in the different fields of science."


Philippe Hernigou
(Paris University, France)

"There are many scientists who can not afford the rather expensive subscriptions to scientific journals. Open access journals offer a good alternative for free access to good quality scientific information."


Fidel Toldrá
(Instituto de Agroquimica y Tecnologia de Alimentos, Spain)

"Open access journals have become a fundamental tool for students, researchers, patients and the general public. Many people from institutions which do not have library or cannot afford to subscribe scientific journals benefit of them on a daily basis. The articles are among the best and cover most scientific areas."


M. Bendandi
(University Clinic of Navarre, Spain)

"These journals provide researchers with a platform for rapid, open access scientific communication. The articles are of high quality and broad scope."


Peter Chiba
(University of Vienna, Austria)

"Open access journals are probably one of the most important contributions to promote and diffuse science worldwide."


Jaime Sampaio
(University of Trás-os-Montes e Alto Douro, Portugal)

"Open access journals make up a new and rather revolutionary way to scientific publication. This option opens several quite interesting possibilities to disseminate openly and freely new knowledge and even to facilitate interpersonal communication among scientists."


Eduardo A. Castro
(INIFTA, Argentina)

"Open access journals are freely available online throughout the world, for you to read, download, copy, distribute, and use. The articles published in the open access journals are high quality and cover a wide range of fields."


Kenji Hashimoto
(Chiba University, Japan)

"Open Access journals offer an innovative and efficient way of publication for academics and professionals in a wide range of disciplines. The papers published are of high quality after rigorous peer review and they are Indexed in: major international databases. I read Open Access journals to keep abreast of the recent development in my field of study."


Daniel Shek
(Chinese University of Hong Kong, Hong Kong)

"It is a modern trend for publishers to establish open access journals. Researchers, faculty members, and students will be greatly benefited by the new journals of Bentham Science Publishers Ltd. in this category."


Jih Ru Hwu
(National Central University, Taiwan)


Browse Contents



Webmaster Contact: info@benthamopen.net
Copyright © 2019 Bentham Open