Table 1.: Automated Antibody Detection Techniques

Vendor System Marker Method/Generation/Format Characteristics

ABBOTT ARCHITECT Antibodies + Antigen Chemiluminescence (CMIA)/4th generation/Sandwich Modular system; 200-800 determinations/hour
Primary tube and aliquot / barcode
25 refrigerated reagents per module / stable 30 days / 5 hours of autonomy
135 samples / module / continuous loading / immediate emergency processing (35 positions/module)
Response time 29 minutes
Sample volume 150 ul / automatic repetitions / conditional tests
Pressure differential clot detection
Intelligent wash system. Minimal carry-over (1 ppm)
Reagents with barcode: Information on batch, expiry date, calibration curves
Long calibration curve stability
Automatic internal quality control system (Levey-Jennings)
Remote diagnostics (AbbottLinkTM).
Connection to Laboratory Information System (LIS)

ROCHE COBAS Antibodies + Antigen Chemiluminescence (ECLIA)/4th generation/Sandwich Modular system; 88-170 determinations/hour
Primary tube and aliquot / barcode
18-25 refrigerated reagents per module
100-150 samples / module / continuous loading / immediate emergency processing
Response time 18 minutes
Sample volume <50 ul /automatic repetitions. Conditional tests
Disposable tips and cups that prevent contamination
Reagents with barcode: Information on batch, expiry date, calibration curves
Calibration stable 28 days.
Connection to Laboratory Information System (LIS)

IZASA ACCESS 2 UnicelTM DxI 400-800 Antibodies + Antigen Chemiluminescence/4th generation/Sandwich Modular system; 200-400 determinations/hour
Primary tube / automatic taking of aliquots and release of primary tube in 5 minutes / barcode
50 refrigerated reagents per module / stable 30 days / autonomy 1200 tests
120 samples / module / continuous loading / immediate emergency processing
Response time 55;
Sample volume <50 ul / automatic repetitions / conditional tests
Fixed tip, no carry-over, wash system with carry-over <10 ppm
Reagents with barcode: Information on batch, expiry date, calibration curves
Calibration curve stability 26 days
Connection to Laboratory Information System (LIS)

DIASORIN LIAISON XL Antibodies + Antigen Separate Chemiluminescence (CLIA)/4th generation/Sandwich 171 determinations/hour
Primary tube / barcode
250 refrigerated reagents per module / stable 8 weeks
120 samples / continuous loading / immediate emergency processing
Response time 40 minutes
Sample volume 350 ul / automatic repetitions / conditional tests
Reagents with barcode: Information on batch, expiry date, calibration curves
Calibration curve stability 15-30 days
Connection to Laboratory Information System (LIS)
Disposable tip / no carry-over / clot detection
Independent antigen and antibody signals / sensitivity of p24 Ag: 1.26 IU/ml

SIEMENS CENTAUROXP and CP
Antibodies Chemiluminescence/3rd generation/Sandwich Disposable tips. Capacity for 840 tips simultaneously
Incubation time 58 min
Sample volume 50 ul
Calibration every 21 days, on-board stability 42 days

Antibodies +Antigen Chemiluminescence/4th generation/Sandwich Disposable tips. Capacity for 840 tips simultaneously
Response time 58 min
Sample volume 100 ul
Calibration every 21 days, on-board stability 42 days
Sensitivity p24 Ag 1.15 IU/ml