The Open Cardiovascular Medicine Journal




ISSN: 1874-1924 ― Volume 13, 2019

Single-Breathhold Four-Dimensional Assessment of Left Ventricular Morphological and Functional Parameters by Magnetic Resonance Imaging Using the VAST Technique



Carlos Eduardo Rochitte1, *, Clerio F Azevedo2, Miguel A Rosário 1, Maria H.R Siqueira1, Victor Monsão1, Manoj Saranathan3, Thomas K Foo3, Roberto Kalil Filho1, Giovanni G Cerri1, José A. F Ramires1
1 Heart Institute (InCor) University of São Paulo Medical School, São Paulo, Brazil
2 D’Or Institute for Research and Education (IDOR), Rio de Janeiro, Brazil
3 GE Healthcare, Baltimore, MD, USA

Abstract

Introduction:

The accurate and reproducible assessment of cardiac volumes, function, and mass is of paramount importance in cardiology. In the present study we sought to determine whether the 3D cine-magnetic resonance (MR) technique, using the variable asymmetric sampling in time (VAST) approach, provided an accurate assessment of LV functional parameters when compared with the conventional 2D cine-MR technique.

Methods:

A total of 43 consecutive patients referred for a CMR examination for clinical reasons and 14 healthy volunteers were included in the study. Cine images were acquired using a steady-state free precession pulse sequence. Two different multiphase acquisitions were performed: conventional 2D cine-MR and 3D cine-MR. The short-axis cine images acquired by both cine-MR techniques were used for the quantitative assessment of LV end-diastolic, end-systolic and stroke volumes, LV mass and ejection fraction.

Results:

All CMR examinations were completed successfully, with both cine-MR imaging techniques yielding interpretable diagnostic results in all patients. Regarding the quantitative assessment, Bland-Altman analyses demonstrated a good agreement between the measurements of both cine-MR techniques for all LV parameters. In addition, the agreement between 2D and 3D cine-MR techniques for the qualitative assessment of LV global function was perfect (kappa = 1.0, P<0.001) for the two observers in consensus. The assessment performed by the third independent observer also demonstrated very good agreement (kappa = 0.88, P<0.001).

Conclusion:

The single breathhold 3D cine-MR technique provides an accurate and reproducible quantitative assessment of LV volumes, mass and function when compared with the conventional 2D cine-MR method.

Keywords: Magnetic resonance imaging, left ventricular function, cine imaging, three-dimensional, steady-state free precession, fast imaging..


Article Information


Identifiers and Pagination:

Year: 2011
Volume: 5
First Page: 90
Last Page: 98
Publisher Id: TOCMJ-5-90
DOI: 10.2174/1874192401105010090

Article History:

Received Date: 28/1/2011
Revision Received Date: 10/2/2011
Acceptance Date: 11/2/2011
Electronic publication date: 22/3/2011
Collection year: 2011

Article Metrics:

CrossRef Citations:
0

Total Statistics:

Full-Text HTML Views: 1085
Abstract HTML Views: 981
PDF Downloads: 270
Total Views/Downloads: 2336

Unique Statistics:

Full-Text HTML Views: 523
Abstract HTML Views: 646
PDF Downloads: 196
Total Views/Downloads: 1365
Geographical View

© Rochitte et al.; Licensee Bentham Open.

open-access license: This is an open access article licensed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/3.0/) which permits unrestricted, non-commercial use, distribution and reproduction in any medium, provided the work is properly cited.


* Address correspondence to this author at Instituto do Coração (InCor) – Setor de Ressonância Magnéti- ca Cardiovascular, Av. Dr. Enéas de Carvalho Aguiar, 44, Andar AB, Cerqueira César, São Paulo, SP, Brazil, 05403-000; Tel: 55-11-3069-5587; Fax: 55-11-3069-5587; E-mail: rochitte@incor.usp.br