PEEK (Polyether-ether-ketone) Based Cervical Total Disc Arthroplasty: Contact Stress and Lubrication Analysis



H Xin, DET Shepherd*, KD Dearn
School of Mechanical Engineering, University of Birmingham, Edgbaston, Birmingham, B15 2TT, UK

Abstract

This paper presents a theoretical analysis of the maximum contact stress and the lubrication regimes for PEEK (Polyether-ether-ketone) based self-mating cervical total disc arthroplasty. The NuNec® cervical disc arthroplasty system was chosen as the study object, which was then analytically modelled as a ball on socket joint. A non-adhesion Hertzian contact model and elastohydrodynamic lubrication theory were used to predict the maximum contact stress and the minimum film thickness, respectively. The peak contact stress and the minimum film thickness between the bearing surfaces were then determined, as the radial clearance or lubricant was varied. The obtained results show that under 150 N loading, the peak contact stress was in the range 5.9 – 32.1 MPa, well below the yield and fatigue strength of PEEK; the calculated minimum film thickness ranged from 0 to 0.042 µm and the corresponding lambda ratio range was from 0 to 0.052. This indicates that the PEEK based cervical disc arthroplasty will operate under a boundary lubrication regime, within the natural angular velocity range of the cervical spine.

Keywords: Cervical disc arthroplasty, contact stress, elastohydrodynamic, hertzain, lubrication, PEEK..


Article Information


Identifiers and Pagination:

Year: 2012
Volume: 6
First Page: 73
Last Page: 79
Publisher Id: TOBEJ-6-73
DOI: 10.2174/1874120701206010073

Article History:

Received Date: 27/1/2012
Revision Received Date: 29/3/2012
Acceptance Date: 29/3/2012
Electronic publication date: 9/5/2012
Collection year: 2012

Article Metrics:

CrossRef Citations:
0

Total Statistics:

Full-Text HTML Views: 828
Abstract HTML Views: 626
PDF Downloads: 224
Total Views/Downloads: 1678

Unique Statistics:

Full-Text HTML Views: 527
Abstract HTML Views: 427
PDF Downloads: 151
Total Views/Downloads: 1105
Geographical View

© Xin et al.; Licensee Bentham Open.

open-access license: This is an open access article licensed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/3.0/) which permits unrestricted, non-commercial use, distribution and reproduction in any medium, provided the work is properly cited.


* Address correspondence to this author at the School of Mechanical Engineering, University of Birmingham, Edgbaston, Birmingham, B15 2TT, UK. Tel: +44 121 414 4266; Fax: +44 121 414 3958; E-mail: d.e.shepherd@bham.ac.uk




1. INTRODUCTION

This paper investigates important design parameters for PEEK (Polyether-ether-ketone) based self-mating cervical total disc arthroplasty. The human spine consists of a series of bony vertebrae that are separated by intervertebral discs, which consist of the inner nucleus pulposus and the outer annulus fibrosus [1Kurtz S M, and Edidin A A, ""The Basic Tools and Terminology of Spine Treatment,"", In: Kurtz S M, and Edidin A A, Eds., Spine Technology Handbook, Academic Press: Burlington, 2006, pp. 1-9.]. The intervertebral disc is susceptible to degenerative diseases, nutritional deficiency, and ageing. A dysfunctional intervertebral disc can lead to spinal pain with a limited degree of motion [2Guerin H A, and Elliott D M, ""Structure and Properties of Soft Tissues in the Spine,"", In: Kurtz S M, and Edidin A A, Eds., Spine Technology Handbook, Academic Press: Burlington, 2006, pp. 35-62.].

There are generally two types of treatments for medium to severe neck pain: fusion surgery (FS) and motion preservation implantation, such as total disc arthroplasty (TDA). FS is considered as the ‘gold standard’ which can successfully relieve pain, however the spinal motion is also removed at the operation level. The primary aim of utilizing TDA is to restore pain-free motion [3Kurtz S M, ""Total Disc Arthroplasty,"", In: Kurtz S M, and Edidin A A, Eds., Spine Technology Handbook, Academic Press: Burlington, 2006, pp. 303-370.]. Cervical TDA is a relatively new technology and there is much debate as to whether its performance is superior to that of fusion [4Murrey D B, Janssen M E, Odum S M, Gottlieb J R, Spector L R, and Darden B V, ""Two-Year Results of a Randomized Controlled Clinical Trial Comparing ProDisc-C and Anterior Cervical Discectomy and Fusion,"", SAS J, vol. 2, pp. 76-85, June 2008.-7Delamarter R B, Murrey D, Janssen M E, Goldstein J A, Zigler J, Tay B K B, and Darden B II, ""Results at 24 months from the prospective, randomized, multicenter Investigational Device Exemption trial of ProDisc-C versus anterior cervical discectomy and fusion with 4-year follow-up and continued access patients,"", SAS J, vol. 4, pp. 122-128, Dec 2010.].

A number of cervical total disc implants that have been approved for use in patients including [3Kurtz S M, ""Total Disc Arthroplasty,"", In: Kurtz S M, and Edidin A A, Eds., Spine Technology Handbook, Academic Press: Burlington, 2006, pp. 303-370.]:

  • Bryan (Medtronic Sofamor Danek, Memphis, TN, USA) – Polyurethane against Titanium bearing coupling;
  • Prestige (Medtronic Sofamor Danek, Memphis, TN, USA) – Stainless steel or Titanium ceramic composite self-mating bearing;
  • Prodisc-C (Synthes, Paoli, PA, USA) – Ultra High Molecular Weight Polyethylene (UHMWPE) against Cobalt Chromium Molybdenum alloy (CoCrMo) bearing coupling;
  • PCM (Cervitech, Rockaway, NJ, USA) – UHMWPE against CoCrMo bearing coupling.

Among these current designs, the ball and socket design configuration dominates. Different material combinations have been used under this configuration to achieve the best tribological and fatigue outcomes; however some degree of inherent deficiencies or defects still exists for these combinations. Metal implants such as the Maverick lumbar disc can release metal ions (cobalt and chromium) in the form of wear debris. This detritus can cause localised tissue ion accumulation and eventually lead to cell death at excessive concentrations [8Hallab N, ""Intervertebral disc joint replacement technology,"", In: Revell P A, Ed., Joint Replacement Technology, Woodhead Publishing Limited: Cambridge, 2008, pp. 515-544., 9Zeh A, Planert M, Siegert G, Lattke P, Held A, and Hein W, " "Release of Cobalt and Chromium Ions Into the Serum Following Implantation of the Metal-on-Metal Maverick-Type Artificial Lumbar Disc (Medtronic Sofamor Danek),"", Spine, vol. 32, pp. 348-352, Feb 2007.]. This is also expected to happen in metal based cervical discs, since wear is an inevitable consequence of normal function. Polymer joint implants (e.g. Ultra High Molecular Weight Polyethylene) generate wear debris which can provoke an adverse biological response (i.e. inflammation), and stimulate osteolysis which finally leads to implant loosening [10Brown S A, ""Synthetic Biomaterials for Spinal Applications,"", Kurtz SM, and Edidin A A, Eds., Spine Technology Handbook, Academic Press: Burlington, pp. 11-33, 2006., 11Ingham E, and Fisher J, ""The role of macrophages in osteolysis of total joint replacement,"", Biomaterials Apr , vol. 26, pp. 1271-1286, Apr 2005.]. With the intention of overcoming the above shortcomings, new bearing materials have been introduced. PEEK (Polyether-ether-ketone) is one such material with great potential for cervical total disc replacement. However, its full performance needs to be analysed to investigate if it overcomes the problems of other material combinations. PEEK is a semi-crystal (i.e. having both amorphous and crystal phases) thermoplastic of the poly-aryl family. The aromatic backbone (i.e. benzene rings) is interconnected by the ether and ketone functional groups. This kind of chemical structure, generally shows high radiation and thermal ageing resistance with good mechanical and tribological performance [12Kurtz S M, and Devine J N, ""PEEK biomaterials in trauma, orthopedic, and spinal implants,"", Biomaterials, vol. 28, pp. 4845-4869, Nov 2007.]. The overall aim of this study was to use theoretical analysis to investigate the maximum contact stress and lubrication regime of PEEK based cervical TDA.

2. MATERIALS AND METHODS

2.1. Model

The NuNec® cervical disc arthroplasty (Pioneer Surgical Technology Inc., Driebergen, Netherlands) was selected as the candidate TDA for this study. The smallest available footprint was chosen with the intention of replicating the worst case operational scenario. This cervical prosthesis consisted of two parts made from PEEK only, and adopted the conventional ball-on-socket design configuration as shown in Fig. (1). A unique cam fixation design is used for immediate short-term fixation, whereby metal blades extend from the device to grip the bone.

Fig. (1)

The NuNec® cervical arthroplasty system a) assembled; b) unassembled.



Fig. (2)

Variation of maximum contact stress with force for radial clearance values of 0.05, 0.10, 0.15 and 0.7 mm.



Fig. (3)

a) Variation of minimum film thickness with angular velocity; b) Variation of Lambda ratio with angular velocity. Each figure is plotted for a cervical disc arthroplasty under 150 N load, using interstitial fluid as the lubricant, at radial clearance values of 0.05, 0.1, 0.15 and 0.7 mm.



Fig. (4)

a) Variation of minimum film thickness with angular velocity; b) Variation of Lambda ratio with angular velocity. Each figure is plotted for a cervical disc arthroplasty at a radial clearance value of 0.7 mm, using interstitial fluid as the lubricant, under three different load levels 0.01, 1 and 150 N.



Fig. (5)

a) Variation of minimum film thickness with angular velocity; b) Variation of Lambda ratio with angular velocity. Each figure is plotted for a cervical disc arthroplasty at a radial clearance value of 0.7 mm, under 150 N load, using two different lubricants bovine serum and interstitial fluid.



The TDA was modelled as a ball-and-socket joint with the ball having a radius R1 and socket having a radius of R2. The Young’s modulus and Poisson’s ratio for both the ball and socket was E and ν, respectively. The radial clearance c between the ball and socket was defined as:


(1) 

2.2. Contact Stress

A Hertzian fully elastic contact model was employed to analyze the maximum superficial contact stress between the bearing surfaces, assumed to be smooth, along with a negligible adhesion force between the contact zone [13Goryacheva IG, ""Contact Mechanics in Tribology,"", , 14Johnson G R, ""Biomechanics of joints,"", In: Revell P A, Ed., Joint Replacement Technology, Woodhead Publishing Limited: Cambridge, 2008, pp. 1-30.]. The maximum contact stress between the bearing surfaces was determined from:


(2) 
where F is the applied force, E' is the equivalent elastic modulus for the two bearing materials and R is the equivalent radius for ball and socket. These two parameters were calculated from:

(3) 
and

(4) 

2.3. Lubrication Regime

The Hamrock-Dowson formula was used to predict the minimum film thickness (hmin) between the bearing surfaces and given as [15Joyce T J, ""Biopolymer Tribology,"", In: Sinha S K, and Briscoe B J, Eds., Polymer Tribology, Imperial College Press: London, 2009, pp. 227-266., 16Mattei L, Di Puccio F, Piccigallo B, and Ciulli E, ""Lubrication and wear modelling of artificial hip joints: a review,"", Tribol Int, vol. 44, pp. 532-549, May 2011.]:


(5) 
where E* is the equivalent modulus of elasticity and uu is the entraining velocity. These two parameters were calculated from:

(6) 
and

(7) 

From these values, the lubrication regime was predicted by calculating the lambda ratio:


(8) 
where σ is the compound surface roughness given by:

(9) 
where Ra1 and Ra2 are the mean surface roughness of the bearing surfaces for the ball and socket, respectively. If the lambda ratio is greater than 3, then a fluid film lubrication regime occurs and complete separation of the bearing surfaces is achieved. A lambda ratio of less than unity indicates boundary lubrication and substantial contact between surface asperities will occur. A lambda ratio of between 1 and 3 indicates a mixed-lubrication regime.

2.4. Parameters

The radii of the ball and socket of the NuNec® cervical disc arthroplasty were measured using a DEA-Swift coordinate measuring machine (CMM) (Hexagon Metrology Ltd., Telford, UK). The principle of the CMM was based on using a 2 mm diameter probe to take 15 randomly selected points across the whole bearing surface of the specimen. The collected data points were then computationally reconstructed to generate the profile of the specimen. Eight repeated measurements were taken for the ball and five repeated measurements were taken for the socket. The socket was found to have a mean radius of 7.010 ± 0.022 mm and the ball had a radius of 6.307 ± 0.004 mm. For analysis the ball radius was chosen as 6.3 mm and the socket radius was chosen as 7 mm, thus giving a radial clearance of 0.7 mm. In this study the aim was to understand the effect of different radial clearances on stress and lubrication regime, so as well as a radial clearance of 0.7 mm, radial clearances of 0.05, 0.1 and 0.15 mm, which have been used by other investigators [17Shaheen A, and Shepherd D E T, ""Lubrication regimes in lumbar total disc arthroplasty,"", Proc. Inst. Mech. Eng. H, vol. 221, pp. 621-627, Jan 2007.-19Wenzel S A, and Shepherd D E T, ""Contact stresses in lumbar total disc arthroplasty,"", Bio-Med Mater. Eng, vol. 17, pp. 169-173, Jan 2007.], were also used in the analysis. Surface roughness of the bearing surfaces was measured using a Taylorsurf-120L (Taylor Hobson Ltd., Leicester, UK) profilometer with a 2 µm radius diamond stylus and an examining area of 5 × 5 mm. Talymap universal 3.1.8 (Taylor Hobson Ltd., Leicester, UK) equipped with a Gaussian filter (cut-off length 0.08 mm) was used to manipulate the data and hence remove measurement noise. The obtained arithmetic mean surface roughness was found to be 0.585 µm for the ball and 0.525 µm for the socket, respectively. The corresponding compound surface roughness value (Eq. 9) was calculated to be 0.786 µm.

The angular velocity for the human cervical spine has been reported to be in the range of 0 to 4.5 rad/s [20Salvia P, Champagne O, Feipel V, Rooze M, and de Beyl D Z, " "Clinical and goniometric evaluation of patients with spasmodic torticollis,"", Clin. Biomech, vol. 21, pp. 323-329, May 2006.]. This can be easily converted into an entraining velocity using Eq. 7. Interstitial fluid was proposed as the natural lubricant for the intervertebral disc, however bovine serum which is normally used for in-vitro wear testing of disc replacements was also considered [21""ISO 18192-1:2011. Implants for surgery - Wear of total intervertebral spinal disc prostheses Part 1: Loading and displacement parameters for wear testing and corresponding environmental conditions for test ISO International Standards,"", ]. An interstitial fluid with a viscosity of 1.24 mPa·s was considered as the primary lubricant [22Windberger U, Bartholovitsch A, Plasenzotti R, Korak K J, and Heinze G, ""Whole blood viscosity, plasma viscosity and erythrocyte aggregation in nine mammalian species: reference values and comparison of data,"", Exp. Physiol, vol. 88, pp. 431-440, May 2003., 23Zurovsky Y, Mitchell G, and Hattingh J, ""Composition and Viscosity of interstitial fluid of rabbits,"", Exp. Physiol, vol. 80, pp. 203-207, Sept 1994.]; 25% bovine serum (diluted with deionised water) with a viscosity of 0.9 mPa·s was used as a comparison [16Mattei L, Di Puccio F, Piccigallo B, and Ciulli E, ""Lubrication and wear modelling of artificial hip joints: a review,"", Tribol Int, vol. 44, pp. 532-549, May 2011., 24Jin Z, Medley J B, and Dowson D, ""Fluid film lubrication in artificial hip joints,"", In: Dowson D, Priest M, Dalmaz G, and Lubrecht A A, Eds., Tribological Research and Design for Engineering Systems, Elsevier: Amsterdam, 2003, pp. 237-256.]. For this study, the maximum compressive loading for the cervical disc was assumed to be 150 N, selected according to the recommended value from the ISO 18192-1 standard for testing intervertebral disc prostheses [21""ISO 18192-1:2011. Implants for surgery - Wear of total intervertebral spinal disc prostheses Part 1: Loading and displacement parameters for wear testing and corresponding environmental conditions for test ISO International Standards,"", ]. The Young’s modulus (E) and Poisson’s ratio (ν) of PEEK were taken to be 3.7 GPa and 0.36, respectively [12Kurtz S M, and Devine J N, ""PEEK biomaterials in trauma, orthopedic, and spinal implants,"", Biomaterials, vol. 28, pp. 4845-4869, Nov 2007., 25""Victrex PEEKTM 450G datasheet," Victrex plc,", Available from http://www.victrex.com/en/victrexlibrary/datasheets/datasheets.php, 2009.]. Finally, the radial clearance was investigated in the range 0.05, 0.1, 0.7 mm and 0.15 mm, while other design parameters were kept constant. A summary of the parameters used are shown in Table 1.

Table 1

Summary of Parameters Used in the Analysis




3. RESULTS

3.1. Contact Stress

Fig. (2) shows the variation of maximum contact stress with force for TDA with different radial clearances. For a PEEK cervical disc under 150 N loading, the maximum contact stress is in the range 5.9 (for a radial clearance of 0.05 mm) to 32.1 MPa (for a radial clearance of 0.7 mm).

3.2. Lubrication Minimum Film Thickness

Fig. (3a & b) show the variation of minimum film thickness and lambda ratio, respectively, by incorporating different radial clearances. The predicted minimum film thickness was in the range 0 to 0.042 µm and the corresponding lambda ratio range was from 0 to 0.052. This indicates that the PEEK cervical discs will operate within a boundary lubrication regime across the whole range of angular velocities, under 150 N loading.

The effects of load level on the minimum film thickness and lambda ratio are displayed in Fig. (4a and 4b), respectively. They show that the loading level has a detrimental effect on both the minimum film thickness and the lambda ratio. Moreover, the obtained lambda ratio is always below unity and indicates boundary lubrication.

Fig. (5a & b) show the variation of minimum film thickness and lambda ratio, respectively, when different lubricants are employed. The minimum film thickness can be enhanced by incorporating a more viscous lubricant; the lambda ratio can be increased in a similar manner. In spite of this, the lubrication regime is still boundary lubrication, regardless of the lubricant used.

4. DISCUSSION

This paper has predicted the likely contact stress and minimum film thickness between the bearing surfaces of a PEEK self-mating cervical disc. The calculated maximum Hertzian contact stress was in the range 5.9 – 32.1 MPa under 150 N loading. Higher stresses were found for discs with a higher radial clearance. These values of stress are well below the yield strength of PEEK 450G under compression which is documented as 120 MPa [25""Victrex PEEKTM 450G datasheet," Victrex plc,", Available from http://www.victrex.com/en/victrexlibrary/datasheets/datasheets.php, 2009.]. Furthermore, the fatigue strength of PEEK 450G with a crystallinity value of 22.5% was reported as 58.72 MPa, at one million cycles [26Tang S M, Cheang P, AbuBakar M S, Khor K A, and Liao K, " "Tension-tension fatigue behavior of hydroxyapatite reinforced polyetheretherketone composites,"", Int. J. Fatigue, vol. 26, pp. 49-57, Jan 2004., 27AbuBakar M S, Cheng M H W, Tang S M, Yu S C, Liao K, Tan C T, Khor K A, and Cheang P, ""Tensile properties tension-tension fatigue and biological response of polyetheretherketone-hydroxyapatite composites for load-bearing orthopedic implants,"", Biomaterials, vol. 24, pp. 2245-2250, June 2003.]. It is predicted that the stress experienced by the contact surface of this disc is insufficient to result in material fatigue. Nevertheless, high cycle fatigue tests of at least 10 million cycles are still required to ensure the structural safety of cervical disc arthroplasty during long-term application [21""ISO 18192-1:2011. Implants for surgery - Wear of total intervertebral spinal disc prostheses Part 1: Loading and displacement parameters for wear testing and corresponding environmental conditions for test ISO International Standards,"", ]. Hertzain contact theory assumes that the contact surface is ideally smooth and frictionless and consequently the interaction force (i.e. adhesion force) between the contact surfaces within the contact zone was not considered. Although other elastic-smooth contact models, such as the Johnson-Kandall-Roberts model take the adhesion force into account [28Nikolai K M, and Alexander V K, ""Adhesion and Friction of Polymers,"", Sujeet K S, and Brian J B, Eds., Polymer tribology, Imperial College Press: London, pp. 3-37, 2009.], there would not be any significant difference between these two models, because the surface energy of PEEK is negligibly small at 0.044 J/m2 [28Nikolai K M, and Alexander V K, ""Adhesion and Friction of Polymers,"", Sujeet K S, and Brian J B, Eds., Polymer tribology, Imperial College Press: London, pp. 3-37, 2009., 29Zhang S, Awaja F, James N, McKenzie D R, and Ruys A J, " "Autohesion of plasma treated semi-crystalline PEEK: Comparative study of argon, nitrogen and oxygen treatments,"", Colloids Surf. A Physicochem Eng. Asp, vol. 374, pp. 88-95, Jan 2011.].

The lubrication analysis assumes steady-state is achieved, under a given constant angular velocity. This is a simplified ideal situation to represent the motion of the neck, which may not be achievable in practice. The intention of adopting this model is to ascertain the potential of cervical discs to operate under a fluid lubrication regime by altering its radial clearance or lubricant. The basic assumption of the elastohydrodynamic theory is that the lubricants (bovine serum and interstitial fluid) utilized are Newtonian fluids and there are no shear-thinning effects [16Mattei L, Di Puccio F, Piccigallo B, and Ciulli E, ""Lubrication and wear modelling of artificial hip joints: a review,"", Tribol Int, vol. 44, pp. 532-549, May 2011.]. According to the obtained lambda ratio results, the PEEK cervical disc, for all parameters, will operate under a boundary lubrication regime, which means that there will be significant surface asperities contact and wear will occur. A previous study has shown that for lumbar disc arthroplasty most devices will also operate with a boundary lubrication regime [17Shaheen A, and Shepherd D E T, ""Lubrication regimes in lumbar total disc arthroplasty,"", Proc. Inst. Mech. Eng. H, vol. 221, pp. 621-627, Jan 2007.]. According to the recent literature, the wear rate of PEEK self-mating cervical disc arthroplasty is reported as 1.4 ± 0.4 mg per million cycles; the historical cervical disc control (i.e. UHMWPE against CoCrMo) has a wear rate of 1.0 ± 0.4 mg per million cycles [30Grupp T M, Meisel H J, Cotton J A, Schwiesau J, Fritz B, Blömer W, and Jansson V, ""Alternative bearing materials for intervertebral disc arthroplasty,"", Biomaterials, vol. 31, pp. 523-531, Jan 2010.].

There are concerns that debris generated as a result of wear may induce an adverse tissue response, such as chronic inflammation and osteolysis. Goodman proposed that the immune response to polymer particles is non-specific and macrophage mediated via the production of cytokines [31Goodman S B, ""Wear particles, periprosthetic osteolysis and the immune system,"", Biomaterials, vol. 28, pp. 5044-5048, Dec 2007.]. It is widely agreed that UHMWPE wear particles in the size range 0.1 to 1.0 µm have been shown to be the most biological reactive [11Ingham E, and Fisher J, ""The role of macrophages in osteolysis of total joint replacement,"", Biomaterials Apr , vol. 26, pp. 1271-1286, Apr 2005., 32Howling G I, Sakoda H, Antonarulrajah A, Marrs H, Stewart T D, Appleyard S, Rand B, Fisher J, and Ingham E, " "Biological response to wear debris generated in carbon based composites as potential bearing surfaces for artificial hip joints,"", J Biomed Mater Res, vol. 67, pp. 758-764, Nov 2003.]. Moreover, quantity, composition, and shape of the wear particles all play important roles in determining the host tissue response [33Utzschneider S, Becker F, Grupp T M, Sievers B, Paulus A, Gottschalk O, and Jansson V, ""Inflammatory response against different carbon fiber-reinforced PEEK wear particles compared with UHMWPE in vivo,"", Acta. Biomater, vol. 6, pp. 4296-4304, Nov 2010.]. There are plenty of studies [32Howling G I, Sakoda H, Antonarulrajah A, Marrs H, Stewart T D, Appleyard S, Rand B, Fisher J, and Ingham E, " "Biological response to wear debris generated in carbon based composites as potential bearing surfaces for artificial hip joints,"", J Biomed Mater Res, vol. 67, pp. 758-764, Nov 2003.-34Rivard CH, Rhalmi S, and Coillard C, ""In vivo biocompatibility testing of PEEK polymer for a spinal implant system: a study in rabbits,"", J. Biomed. Mater. Res, vol. 62, pp. 488-498, Dec 2002.] that report on the biocompatibility of PEEK wear debris, however none focus on wear debris generated in cervical TDA. A recent in-vitro wear study pointed out that the wear particle size distribution of a PEEK-on-PEEK cervical disc was from 0.23 to 0.51 µm with a mean roundness of 0.5 [30Grupp T M, Meisel H J, Cotton J A, Schwiesau J, Fritz B, Blömer W, and Jansson V, ""Alternative bearing materials for intervertebral disc arthroplasty,"", Biomaterials, vol. 31, pp. 523-531, Jan 2010.]. A more targeted host tissue response study is demanded to further ascertain the biocompatibility of PEEK wear debris generated from cervical total disc replacement.

Cervical disc arthroplasty is a relatively new technology and there is still much debate as to whether its performance is actually superior to that of fusion [4Murrey D B, Janssen M E, Odum S M, Gottlieb J R, Spector L R, and Darden B V, ""Two-Year Results of a Randomized Controlled Clinical Trial Comparing ProDisc-C and Anterior Cervical Discectomy and Fusion,"", SAS J, vol. 2, pp. 76-85, June 2008.-7Delamarter R B, Murrey D, Janssen M E, Goldstein J A, Zigler J, Tay B K B, and Darden B II, ""Results at 24 months from the prospective, randomized, multicenter Investigational Device Exemption trial of ProDisc-C versus anterior cervical discectomy and fusion with 4-year follow-up and continued access patients,"", SAS J, vol. 4, pp. 122-128, Dec 2010.]. This study has discussed a new bearing material combination of PEEK against PEEK, but other devices have a longer record of use. Currently, there are three cervical disc arthroplasty (Prestige ST, Bryan, and ProDisc-C) undergoing a US FDA investigational device exemption trial [35Upadhyaya C D, Wu J C, Trost G, Haid R W, Traynelis V C, Tay B, Coric D, and Mummaneni P V, "“Analysis of the three United States Food and Drug Administration investigational device exemptioncervical arthroplasty trials Clinical article,”", J. Neurosurg. Spine, vol. 16, pp. 216-228, Mar 2012.]. Studies that have investigated the wear of other cervical disc arthroplasty are few. Anderson et al. [36Anderson PA, Rouleau J P, Bryan VE, and Carlson CS, "“Wear analysis of the Bryan cervical disc prosthesis,”", Spine, vol. 28, pp. S186-S194, Oct 2003.] found that the Bryan disc had a wear rate of 1.2 mg per million cycles and that this was satisfactory. Implantation of of a porous coated cervical disc arthroplasty into goats showed no inflammatory response from wear debris after six months implanted in the animals [37McAfee P C, Cunningham B, Dmitriev A, Hu N B, Kim S W, Cappuccino A, and Pimenta L, "“Cervical disc replacement - porous coated motion prosthesis - a comparative biomechanical analysis showing the key role of the posterior longitudinal ligament,”", Spine, vol. 28, pp. S176-S185, Oct 2003.]. Lumbar disc arthroplasty have been studied in more detail [38Moghadas P, Mahomed A, Hukins D W L, and Shepherd D E T, " “Friction in metal-on-metal total disc arthroplasty: effect of ball radius,”", J. Biomech, vol. 45, pp. 504-509, Feb 2012.-40Pare P E, Chan F W, and Powell M L, "“Wear characterization of the A-MAV (TM) anterior motion replacement using a spine wear simulator,”", Wear, vol. 263, pp. 1055-1059, Sep 2007.], but a direct comparison with cervical discs is unwise due to the different motions and loads.

CONCLUSIONS

From this theoretical study, it can be confirmed that a PEEK based artificial cervical disc will operate under a boundary lubrication regime, with the generation of wear debris likely. The maximum contact stress was found to be 32.1 MPa, but this is less than either the compressive yield strength or fatigue strength of PEEK 450G.

CONFLICT OF INTEREST

None declared.

ACKNOWLEDGEMENTS

We thank Dr William A. Murray for the help with operating the Taylorsurf-120L and DEA-Swift equipment.

GLOSSARY

R1  = Radius curvature of ball component.
R2  = Radius curvature of socket component.
c  = Radial clearance equals to the difference between the R2 and R1.
Pmax  = Maximum contact stress.
F  = Applied force.
E´  = Equivalent elastic modulus for the two bearing material.
R  = Equivalent radius for ball and socket.
E  = Youth’s modulus of the material.
ν  = Poisson’s ratio of the material.
hmin  = Minimum film thickness.
µ  = Viscosity of the lubricant.
u  = Entrance velocity of the disc.
E*  = Equivalent modulus of elasticity.
λ  = Lambda ratio.
σ  = Compound surface roughness.

REFERENCES

[1] Kurtz S M, and Edidin A A, ""The Basic Tools and Terminology of Spine Treatment,"", In: Kurtz S M, and Edidin A A, Eds., Spine Technology Handbook, Academic Press: Burlington, 2006, pp. 1-9.
[2] Guerin H A, and Elliott D M, ""Structure and Properties of Soft Tissues in the Spine,"", In: Kurtz S M, and Edidin A A, Eds., Spine Technology Handbook, Academic Press: Burlington, 2006, pp. 35-62.
[3] Kurtz S M, ""Total Disc Arthroplasty,"", In: Kurtz S M, and Edidin A A, Eds., Spine Technology Handbook, Academic Press: Burlington, 2006, pp. 303-370.
[4] Murrey D B, Janssen M E, Odum S M, Gottlieb J R, Spector L R, and Darden B V, ""Two-Year Results of a Randomized Controlled Clinical Trial Comparing ProDisc-C and Anterior Cervical Discectomy and Fusion,"", SAS J, vol. 2, pp. 76-85, June 2008.
[5] Hilibrand A S, and Robbins M, ""Adjacent segment degeneration and adjacent segment disease: the consequences of spinal fusion?,"", Spine J, vol. 4, pp. S190-S194, Nov 2004.
[6] Robertson J T, Papadopoulo S M, and Traynelis V C, " "Assessment of adjacent-segment disease in patients treated with cervical fusion or arthroplasty: a prospective 2-year study,"", J. Neurosurg Spine, vol. 3, pp. 417-423, Dec 2005.
[7] Delamarter R B, Murrey D, Janssen M E, Goldstein J A, Zigler J, Tay B K B, and Darden B II, ""Results at 24 months from the prospective, randomized, multicenter Investigational Device Exemption trial of ProDisc-C versus anterior cervical discectomy and fusion with 4-year follow-up and continued access patients,"", SAS J, vol. 4, pp. 122-128, Dec 2010.
[8] Hallab N, ""Intervertebral disc joint replacement technology,"", In: Revell P A, Ed., Joint Replacement Technology, Woodhead Publishing Limited: Cambridge, 2008, pp. 515-544.
[9] Zeh A, Planert M, Siegert G, Lattke P, Held A, and Hein W, " "Release of Cobalt and Chromium Ions Into the Serum Following Implantation of the Metal-on-Metal Maverick-Type Artificial Lumbar Disc (Medtronic Sofamor Danek),"", Spine, vol. 32, pp. 348-352, Feb 2007.
[10] Brown S A, ""Synthetic Biomaterials for Spinal Applications,"", Kurtz SM, and Edidin A A, Eds., Spine Technology Handbook, Academic Press: Burlington, pp. 11-33, 2006.
[11] Ingham E, and Fisher J, ""The role of macrophages in osteolysis of total joint replacement,"", Biomaterials Apr , vol. 26, pp. 1271-1286, Apr 2005.
[12] Kurtz S M, and Devine J N, ""PEEK biomaterials in trauma, orthopedic, and spinal implants,"", Biomaterials, vol. 28, pp. 4845-4869, Nov 2007.
[13] Goryacheva IG, ""Contact Mechanics in Tribology,"",
[14] Johnson G R, ""Biomechanics of joints,"", In: Revell P A, Ed., Joint Replacement Technology, Woodhead Publishing Limited: Cambridge, 2008, pp. 1-30.
[15] Joyce T J, ""Biopolymer Tribology,"", In: Sinha S K, and Briscoe B J, Eds., Polymer Tribology, Imperial College Press: London, 2009, pp. 227-266.
[16] Mattei L, Di Puccio F, Piccigallo B, and Ciulli E, ""Lubrication and wear modelling of artificial hip joints: a review,"", Tribol Int, vol. 44, pp. 532-549, May 2011.
[17] Shaheen A, and Shepherd D E T, ""Lubrication regimes in lumbar total disc arthroplasty,"", Proc. Inst. Mech. Eng. H, vol. 221, pp. 621-627, Jan 2007.
[18] Clewlow J P, Pylios T, and Shepherd D E T, ""Soft layer bearing joints for spine arthroplasty,"", Mater Des, vol. 29, pp. 1981-1985, Dec 2008.
[19] Wenzel S A, and Shepherd D E T, ""Contact stresses in lumbar total disc arthroplasty,"", Bio-Med Mater. Eng, vol. 17, pp. 169-173, Jan 2007.
[20] Salvia P, Champagne O, Feipel V, Rooze M, and de Beyl D Z, " "Clinical and goniometric evaluation of patients with spasmodic torticollis,"", Clin. Biomech, vol. 21, pp. 323-329, May 2006.
[21] ""ISO 18192-1:2011. Implants for surgery - Wear of total intervertebral spinal disc prostheses Part 1: Loading and displacement parameters for wear testing and corresponding environmental conditions for test ISO International Standards,"",
[22] Windberger U, Bartholovitsch A, Plasenzotti R, Korak K J, and Heinze G, ""Whole blood viscosity, plasma viscosity and erythrocyte aggregation in nine mammalian species: reference values and comparison of data,"", Exp. Physiol, vol. 88, pp. 431-440, May 2003.
[23] Zurovsky Y, Mitchell G, and Hattingh J, ""Composition and Viscosity of interstitial fluid of rabbits,"", Exp. Physiol, vol. 80, pp. 203-207, Sept 1994.
[24] Jin Z, Medley J B, and Dowson D, ""Fluid film lubrication in artificial hip joints,"", In: Dowson D, Priest M, Dalmaz G, and Lubrecht A A, Eds., Tribological Research and Design for Engineering Systems, Elsevier: Amsterdam, 2003, pp. 237-256.
[25] ""Victrex PEEKTM 450G datasheet," Victrex plc,", Available from http://www.victrex.com/en/victrexlibrary/datasheets/datasheets.php, 2009.
[26] Tang S M, Cheang P, AbuBakar M S, Khor K A, and Liao K, " "Tension-tension fatigue behavior of hydroxyapatite reinforced polyetheretherketone composites,"", Int. J. Fatigue, vol. 26, pp. 49-57, Jan 2004.
[27] AbuBakar M S, Cheng M H W, Tang S M, Yu S C, Liao K, Tan C T, Khor K A, and Cheang P, ""Tensile properties tension-tension fatigue and biological response of polyetheretherketone-hydroxyapatite composites for load-bearing orthopedic implants,"", Biomaterials, vol. 24, pp. 2245-2250, June 2003.
[28] Nikolai K M, and Alexander V K, ""Adhesion and Friction of Polymers,"", Sujeet K S, and Brian J B, Eds., Polymer tribology, Imperial College Press: London, pp. 3-37, 2009.
[29] Zhang S, Awaja F, James N, McKenzie D R, and Ruys A J, " "Autohesion of plasma treated semi-crystalline PEEK: Comparative study of argon, nitrogen and oxygen treatments,"", Colloids Surf. A Physicochem Eng. Asp, vol. 374, pp. 88-95, Jan 2011.
[30] Grupp T M, Meisel H J, Cotton J A, Schwiesau J, Fritz B, Blömer W, and Jansson V, ""Alternative bearing materials for intervertebral disc arthroplasty,"", Biomaterials, vol. 31, pp. 523-531, Jan 2010.
[31] Goodman S B, ""Wear particles, periprosthetic osteolysis and the immune system,"", Biomaterials, vol. 28, pp. 5044-5048, Dec 2007.
[32] Howling G I, Sakoda H, Antonarulrajah A, Marrs H, Stewart T D, Appleyard S, Rand B, Fisher J, and Ingham E, " "Biological response to wear debris generated in carbon based composites as potential bearing surfaces for artificial hip joints,"", J Biomed Mater Res, vol. 67, pp. 758-764, Nov 2003.
[33] Utzschneider S, Becker F, Grupp T M, Sievers B, Paulus A, Gottschalk O, and Jansson V, ""Inflammatory response against different carbon fiber-reinforced PEEK wear particles compared with UHMWPE in vivo,"", Acta. Biomater, vol. 6, pp. 4296-4304, Nov 2010.
[34] Rivard CH, Rhalmi S, and Coillard C, ""In vivo biocompatibility testing of PEEK polymer for a spinal implant system: a study in rabbits,"", J. Biomed. Mater. Res, vol. 62, pp. 488-498, Dec 2002.
[35] Upadhyaya C D, Wu J C, Trost G, Haid R W, Traynelis V C, Tay B, Coric D, and Mummaneni P V, "“Analysis of the three United States Food and Drug Administration investigational device exemptioncervical arthroplasty trials Clinical article,”", J. Neurosurg. Spine, vol. 16, pp. 216-228, Mar 2012.
[36] Anderson PA, Rouleau J P, Bryan VE, and Carlson CS, "“Wear analysis of the Bryan cervical disc prosthesis,”", Spine, vol. 28, pp. S186-S194, Oct 2003.
[37] McAfee P C, Cunningham B, Dmitriev A, Hu N B, Kim S W, Cappuccino A, and Pimenta L, "“Cervical disc replacement - porous coated motion prosthesis - a comparative biomechanical analysis showing the key role of the posterior longitudinal ligament,”", Spine, vol. 28, pp. S176-S185, Oct 2003.
[38] Moghadas P, Mahomed A, Hukins D W L, and Shepherd D E T, " “Friction in metal-on-metal total disc arthroplasty: effect of ball radius,”", J. Biomech, vol. 45, pp. 504-509, Feb 2012.
[39] Vicars R, Hyde P J, Brown T D, Tipper J L, Ingham E, Fisher J, and Hall R M, "“The effect of anterior-posterior shear load on the wear of ProDisc-L TDR,”", Eur. Spine J, vol. 19, pp. 1356-1362, Aug 2010.
[40] Pare P E, Chan F W, and Powell M L, "“Wear characterization of the A-MAV (TM) anterior motion replacement using a spine wear simulator,”", Wear, vol. 263, pp. 1055-1059, Sep 2007.

Endorsements



"Open access will revolutionize 21st century knowledge work and accelerate the diffusion of ideas and evidence that support just in time learning and the evolution of thinking in a number of disciplines."


Daniel Pesut
(Indiana University School of Nursing, USA)

"It is important that students and researchers from all over the world can have easy access to relevant, high-standard and timely scientific information. This is exactly what Open Access Journals provide and this is the reason why I support this endeavor."


Jacques Descotes
(Centre Antipoison-Centre de Pharmacovigilance, France)

"Publishing research articles is the key for future scientific progress. Open Access publishing is therefore of utmost importance for wider dissemination of information, and will help serving the best interest of the scientific community."


Patrice Talaga
(UCB S.A., Belgium)

"Open access journals are a novel concept in the medical literature. They offer accessible information to a wide variety of individuals, including physicians, medical students, clinical investigators, and the general public. They are an outstanding source of medical and scientific information."


Jeffrey M. Weinberg
(St. Luke's-Roosevelt Hospital Center, USA)

"Open access journals are extremely useful for graduate students, investigators and all other interested persons to read important scientific articles and subscribe scientific journals. Indeed, the research articles span a wide range of area and of high quality. This is specially a must for researchers belonging to institutions with limited library facility and funding to subscribe scientific journals."


Debomoy K. Lahiri
(Indiana University School of Medicine, USA)

"Open access journals represent a major break-through in publishing. They provide easy access to the latest research on a wide variety of issues. Relevant and timely articles are made available in a fraction of the time taken by more conventional publishers. Articles are of uniformly high quality and written by the world's leading authorities."


Robert Looney
(Naval Postgraduate School, USA)

"Open access journals have transformed the way scientific data is published and disseminated: particularly, whilst ensuring a high quality standard and transparency in the editorial process, they have increased the access to the scientific literature by those researchers that have limited library support or that are working on small budgets."


Richard Reithinger
(Westat, USA)

"Not only do open access journals greatly improve the access to high quality information for scientists in the developing world, it also provides extra exposure for our papers."


J. Ferwerda
(University of Oxford, UK)

"Open Access 'Chemistry' Journals allow the dissemination of knowledge at your finger tips without paying for the scientific content."


Sean L. Kitson
(Almac Sciences, Northern Ireland)

"In principle, all scientific journals should have open access, as should be science itself. Open access journals are very helpful for students, researchers and the general public including people from institutions which do not have library or cannot afford to subscribe scientific journals. The articles are high standard and cover a wide area."


Hubert Wolterbeek
(Delft University of Technology, The Netherlands)

"The widest possible diffusion of information is critical for the advancement of science. In this perspective, open access journals are instrumental in fostering researches and achievements."


Alessandro Laviano
(Sapienza - University of Rome, Italy)

"Open access journals are very useful for all scientists as they can have quick information in the different fields of science."


Philippe Hernigou
(Paris University, France)

"There are many scientists who can not afford the rather expensive subscriptions to scientific journals. Open access journals offer a good alternative for free access to good quality scientific information."


Fidel Toldrá
(Instituto de Agroquimica y Tecnologia de Alimentos, Spain)

"Open access journals have become a fundamental tool for students, researchers, patients and the general public. Many people from institutions which do not have library or cannot afford to subscribe scientific journals benefit of them on a daily basis. The articles are among the best and cover most scientific areas."


M. Bendandi
(University Clinic of Navarre, Spain)

"These journals provide researchers with a platform for rapid, open access scientific communication. The articles are of high quality and broad scope."


Peter Chiba
(University of Vienna, Austria)

"Open access journals are probably one of the most important contributions to promote and diffuse science worldwide."


Jaime Sampaio
(University of Trás-os-Montes e Alto Douro, Portugal)

"Open access journals make up a new and rather revolutionary way to scientific publication. This option opens several quite interesting possibilities to disseminate openly and freely new knowledge and even to facilitate interpersonal communication among scientists."


Eduardo A. Castro
(INIFTA, Argentina)

"Open access journals are freely available online throughout the world, for you to read, download, copy, distribute, and use. The articles published in the open access journals are high quality and cover a wide range of fields."


Kenji Hashimoto
(Chiba University, Japan)

"Open Access journals offer an innovative and efficient way of publication for academics and professionals in a wide range of disciplines. The papers published are of high quality after rigorous peer review and they are Indexed in: major international databases. I read Open Access journals to keep abreast of the recent development in my field of study."


Daniel Shek
(Chinese University of Hong Kong, Hong Kong)

"It is a modern trend for publishers to establish open access journals. Researchers, faculty members, and students will be greatly benefited by the new journals of Bentham Science Publishers Ltd. in this category."


Jih Ru Hwu
(National Central University, Taiwan)


Browse Contents



Webmaster Contact: info@benthamopen.net
Copyright © 2019 Bentham Open