Successful Closed Reduction of a Dislocated Constrained Total Hip Arthroplasty: A Case Report and Literature Review



Motoki Sonohata*, 1, 2, Warit Waewsawangwong 1, 3, Stuart B Goodman 1
1 Department of Orthopaedic Surgery, Stanford School of Medicine, Stanford University, Stanford, California, USA
2 Department of Orthopedic Surgery, Faculty of Medicine, Saga University, Saga, Japan
3 Department of Orthopedic Surgery, Somdej Prapinkao Hospital, Bangkok, Thailand

Abstract

Many surgeons use acetabular components with constrained polyethylene liners to improve stability after total hip arthroplasty in patients with a history of hip dislocation. Surgical treatment is generally thought to be the only available option for the dislocated constrained liner. The success rate and clinical results of closed reduction for such patients is unclear. This report presents a case of a successful closed reduction of a dislocated constrained liner. Few papers have so far addressed closed reduction of a dislocated constrained liner. Furthermore, previous studies reported a variety of components. Publication of additional successful and unsuccessful case reports is therefore needed to help establish the optimal treatment protocol for a dislocated constrained liner.

Keywords: Closed reduction, dislocation, acetabular, constrained liner, total hip arthroplasty, complication.


Article Information


Identifiers and Pagination:

Year: 2012
Volume: 6
First Page: 211
Last Page: 214
Publisher Id: TOORTHJ-6-211
DOI: 10.2174/1874325001206010211

Article History:

Received Date: 26/1/2012
Revision Received Date: 2/5/2012
Acceptance Date: 4/5/2012
Electronic publication date: 30/5/2012
Collection year: 2012

Article Metrics:

CrossRef Citations:
0

Total Statistics:

Full-Text HTML Views: 1852
Abstract HTML Views: 925
PDF Downloads: 272
Total Views/Downloads: 3049

Unique Statistics:

Full-Text HTML Views: 928
Abstract HTML Views: 523
PDF Downloads: 175
Total Views/Downloads: 1626
Geographical View

© Sonohata et al.; Licensee Bentham Open.

open-access license: This is an open access article licensed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/3.0/) which permits unrestricted, non-commercial use, distribution and reproduction in any medium, provided the work is properly cited.


* Address correspondence to this author at the Department of Orthopaedic Surgery, Faculty of Medicine, Saga University, Saga 849-8501, Japan; Tel: +81-952-34-2343; Fax: +81-952-34-2059; E-mail: epc9719@yahoo.co.jp




INTRODUCTION

Dislocation after total hip arthroplasty (THA) ranges from less than 1% to 6% of primary cases [1 Soong M, Rubash HE, Macaulay W. Dislocation after total hip arthroplasty J Am Acad Orthop Surg 2004; 12(5): 314-21.] and from 15% to 30% of revision cases [2 Su EP, Pellicci PM. The role of constrained liners in total hip arthroplasty Clin Orthop 2004; 420: 122-9.]. Unfortunately, the success rate of nonoperative treatment of dislocation after THA can be unreliable, and a third of such patients have recurrences1. Surgical procedures used to treat instability and dislocation include tightening the abductor musculature, removing sources of impingement, repositioning malaligned components, and using acetabular liners with elevated rims, however, such treatments fail in 30% to 50% of patients [3 Morrey BF. Difficult complications after hip joint replacement Dislocation Clin Orthop 1997; 344: 179-87.]. The use of constrained liners, which relies on a locking mechanism to capture the femoral head, has been developed to help manage this problem [4 Shrader MW, Parvizi J, Lewallen DG. The use of a constrained acetabular component to treat instability after total hip arthroplasty J Bone Joint Surg Am 2003; 85(11): 2179-83.]. Despite such efforts, patients continue to be at high risk for instability and 3% to 18% experience recurrent dislocation even after constrained components have been implanted [5 Berend KR, Lombardi AV Jr, Mallory TH, Adams JB, Russell JH, Groseth KL. The long-term outcome of 755 consecutive constrained acetabular components in total hip arthroplasty examining the successes and failures J Arthroplasty 2005; 20(7 Suppl. 3): 93-102., 6 Khan RJ, Fick D, Alakeson R, et al. A constrained acetabular component for recurrent dislocation J Bone Joint Surg Br 2006; 88(7): 870-6.]. Generally, open surgical reduction is thought to be the only available treatment for such cases, however, this compromises the integrity of the joint and exposes the patient to the additional risk of surgery [7 McPherson EJ, Costigan WM, Gerhardt MB, Norris LR. Closed reduction of dislocated total hip with S-ROM constrained acetabular component J Arthroplasty 1999; 14(7): 882-5.]. Several authors have reported closed reduction for dislocation of a constrained liner [7 McPherson EJ, Costigan WM, Gerhardt MB, Norris LR. Closed reduction of dislocated total hip with S-ROM constrained acetabular component J Arthroplasty 1999; 14(7): 882-5.-13 Flint JH, Phisitkul P, Callaghan JJ. Closed reduction of a dislocated constrained total hip arthroplasty using a novel technique with a peg board Orthopedics 2010; 10: 201-3.]. However, constrained acetabular liners are currently available in various designs, with differences in the locking mechanisms. To our knowledge, this is the first reported case of a successful closed reduction of a dislocated constrained THA using a Trilogy Acetabular System Constrained liner (Zimmer, Inc, Warsaw, Indiana, USA).

The study protocol adhered to the ethics guidelines of the 1975 Declaration of Helsinki, and the study was approved by the Institutional Review Board of Stanford University.

CASE

A 79-year-old female underwent primary THA for nonunion of a right subtrochanteric fracture of the femur with degenerative arthritis of the right hip joint using an uncemented Trilogy acetabular system and fully porous-coated stem (Zimmer, Warsaw, Indiana, USA). Five years later, at age 84, the hip was revised to a Trilogy constrained liner with a 10-degree oblique face and 32-mm head (Zimmer, Inc, Warsaw, Indiana, USA) (Fig. 1) for instability and recurrent dislocation (Fig. 2). However, 5 years after the revision surgery, at age 88, she bent over, hyperflexed the hip and complained of right hip pain. A radiograph in the emergency department in our hospital demonstrated dislocation of the constrained THA. The head appeared to have dislocated from the liner, and the locking ring disengaged but was not broken (Fig. 3). A closed reduction was performed in the emergency department with sedation. The head was relocated and the locking ring was repositioned to where it was previously located (Fig. 4). The patient’s hip was placed in an abduction brace with full weight bearing. Further radiographs demonstrated no further change. The patient could ambulate with minimal aid. The patient died 10 months later of unrelated causes.

Fig. (1)

The photograph shows the Trilogy constrained liner with a 10-degree oblique face (Zimmer, Inc, Warsaw, Indiana, USA).



Fig. (2)

Radiographs before dislocation. (A) Anteroposterior view, (B) Lateral view.



Fig. (3)

Anteroposterior radiograph of dislocation of the right hip.



Fig. (4)

Radiographs after closed reduction. (A) Anteroposterior view, (B) Lateral view.



DISCUSSION

Dislocation after THA using a constrained polyethylene liner presents a substantial treatment challenge. Closed reduction provides a conservative treatment option that avoids the risk of open surgical reduction. This point is especially relevant considering the surgical complications that exist in older patients with recurrent instability and a complex revision history [9 Harman MK, Hodge WA, Banks SA. Closed reduction of constrained total hip arthroplasty Clin Orthop 2003; 414: 121-8.].

Several authors have described successful closed reduction of a dislocated constrained liner [7 McPherson EJ, Costigan WM, Gerhardt MB, Norris LR. Closed reduction of dislocated total hip with S-ROM constrained acetabular component J Arthroplasty 1999; 14(7): 882-5.-13 Flint JH, Phisitkul P, Callaghan JJ. Closed reduction of a dislocated constrained total hip arthroplasty using a novel technique with a peg board Orthopedics 2010; 10: 201-3.]. Generally, it is thought that plastic deformation must occur in the polyethylene during the initial dislocation and during the closed reduction process. The torque required for subsequent dislocations for a second and third time decreased by 24% and 8% in the Poly-Dial constrained liner (DePuy, Warsaw, Indiana, USA), respectively [9 Harman MK, Hodge WA, Banks SA. Closed reduction of constrained total hip arthroplasty Clin Orthop 2003; 414: 121-8.]. However, not all patients undergoing closed reduction of constrained liners require revision surgery or experience recurrent dislocations (see Table 1).

Table 1

Summary of Literature Review on Closed Reduction of Constrained Liner




Harman et al. [9 Harman MK, Hodge WA, Banks SA. Closed reduction of constrained total hip arthroplasty Clin Orthop 2003; 414: 121-8.] reported that 6 hips remained stable 7 to 72 months after the last reduction. They noted that 2 of 6 hips underwent repeated successful closed reduction within 6 months of their initial dislocation and reduction and remained stable for at least 3 years. In contrast, McPherson et al. [7 McPherson EJ, Costigan WM, Gerhardt MB, Norris LR. Closed reduction of dislocated total hip with S-ROM constrained acetabular component J Arthroplasty 1999; 14(7): 882-5.] reported that 3 hips with constrained liners required further operation after closed reduction. Two patients underwent revision surgery after 1 and 3 months their initial dislocation and reduction, and one patient was converted to a Girdlestone resection arthroplasty after 6 months. Our current case underwent closed reduction once, and did not require any further surgery.

The closed reduction techniques of dislocated constrained liner were reported in 7 papers [7 McPherson EJ, Costigan WM, Gerhardt MB, Norris LR. Closed reduction of dislocated total hip with S-ROM constrained acetabular component J Arthroplasty 1999; 14(7): 882-5.-13 Flint JH, Phisitkul P, Callaghan JJ. Closed reduction of a dislocated constrained total hip arthroplasty using a novel technique with a peg board Orthopedics 2010; 10: 201-3.]. Adequate anesthesia and usage of fluoroscopy were common described in these papers. The procedure of closed reduction for dislocated liner was formed in 2 steps. First, traction was applied and the femoral head was placed in a “perched” position on the acetabular cup. The hip was positioned in 10-40 degrees abduction during traction [7 McPherson EJ, Costigan WM, Gerhardt MB, Norris LR. Closed reduction of dislocated total hip with S-ROM constrained acetabular component J Arthroplasty 1999; 14(7): 882-5., 9 Harman MK, Hodge WA, Banks SA. Closed reduction of constrained total hip arthroplasty Clin Orthop 2003; 414: 121-8.-12 Gaines RJ, Hardenbrook M. Closed reduction of a dislocated total hip arthroplasty with a constrained acetabular component Am J Orthop (Belle Mead NJ) 2009; 38(10): 523-.]. Additional hip flexion was described in 2 papers, in full extension in one [9 Harman MK, Hodge WA, Banks SA. Closed reduction of constrained total hip arthroplasty Clin Orthop 2003; 414: 121-8.], however, Flint et al. adopted in-line traction for the limb [13 Flint JH, Phisitkul P, Callaghan JJ. Closed reduction of a dislocated constrained total hip arthroplasty using a novel technique with a peg board Orthopedics 2010; 10: 201-3.].

Second, the femoral head should be passed through the constrained liner with axial compression after confirmation that the femoral head was perched just lateral to the rim. In this phase, hip flexion was increased to 30-90 degrees from first step [7 McPherson EJ, Costigan WM, Gerhardt MB, Norris LR. Closed reduction of dislocated total hip with S-ROM constrained acetabular component J Arthroplasty 1999; 14(7): 882-5., 9 Harman MK, Hodge WA, Banks SA. Closed reduction of constrained total hip arthroplasty Clin Orthop 2003; 414: 121-8., 11 Robertson WJ, Mattern CJ, Hur J, Su EP, Pellicci PM. Failure mechanisms and closed reduction of a constrained tripolar acetabular liner J Arthroplast 2009; 24(2): 322.e5-11.].

Direct medial pressure was also placed to the greater trochanter [7 McPherson EJ, Costigan WM, Gerhardt MB, Norris LR. Closed reduction of dislocated total hip with S-ROM constrained acetabular component J Arthroplasty 1999; 14(7): 882-5., 9 Harman MK, Hodge WA, Banks SA. Closed reduction of constrained total hip arthroplasty Clin Orthop 2003; 414: 121-8., 10 Birdwell S, Wilke E. Closed reduction of constrained total hip arthroplasty in the ED J Emerg Med 2011; 40(2): 162-6., 13 Flint JH, Phisitkul P, Callaghan JJ. Closed reduction of a dislocated constrained total hip arthroplasty using a novel technique with a peg board Orthopedics 2010; 10: 201-3.]. Gaines et al. [12 Gaines RJ, Hardenbrook M. Closed reduction of a dislocated total hip arthroplasty with a constrained acetabular component Am J Orthop (Belle Mead NJ) 2009; 38(10): 523-.] reported an anomalous closed reduction. They initially failed at closed reduction, however, the patient’s femoral head subsequently reduced naturally the next day.

The most commonly used femoral heads for constrained liners are 28 mm and 32 mm in diameter. Harman et al. [9 Harman MK, Hodge WA, Banks SA. Closed reduction of constrained total hip arthroplasty Clin Orthop 2003; 414: 121-8.] reported no successful closed reductions of a 32-mm femoral head was achieved in their 8 patients. They described that the force to relocate the 28-mm heads into constrained liners in vitro was 1380 N (310 Ib), whereas 32-mm heads required greater force. Six 28-mm and one 32-mm head were relocated into constrained liners [9 Harman MK, Hodge WA, Banks SA. Closed reduction of constrained total hip arthroplasty Clin Orthop 2003; 414: 121-8.-13 Flint JH, Phisitkul P, Callaghan JJ. Closed reduction of a dislocated constrained total hip arthroplasty using a novel technique with a peg board Orthopedics 2010; 10: 201-3.], however, the details of implants and clinical results at follow-up concerning of the case of 32-mm head were unclear [10 Birdwell S, Wilke E. Closed reduction of constrained total hip arthroplasty in the ED J Emerg Med 2011; 40(2): 162-6.]. In the current case, the patient’ femoral head was 32- mm head with Trilogy constrained liner (Zimmer, Warsaw, Indiana, USA).

Four constrained liners are currently in common use: Omnifit liner (Stryker, Mahwah, New Jersey, USA), S-ROM liner/Poly-Dial (Depuy, Warsaw, Indiana, USA), RingLoc constrained acetabular liner (Biomet, Warsaw, Indiana, USA), and Trilogy constrained liner (Zimmer, Warsaw, Indiana, USA). Given substantial differences among constrained components from different implant manufactures, several reported closed reduction techniques may not be applicable to specific constrained acetabular components. This case is the first report of successful closed reduction of a dislocated constrained total hip arthroplasty using the Trilogy constrained liner and a 32 mm head.

In the aging population with complex medical issues, closed reduction may obviate the need for a more invasive open procedure or at least restore patient function until revision THA can be electivity planned. An attempt at closed reduction may be indicated unless the acetabular component has failed at the liner-shell interface, shell-bone interface, or the locking ring has fractured.

The limitation of this case report was that the follow-up period was only 10 months. Therefore, it is unclear whether recurrences of dislocation might happen with longer term follow-up.

CONCLUSION

We describe a case of successful closed reduction of a dislocated Trilogy constrained liner (Zimmer, Warsaw, Indiana, USA). After the reduction, recurrent dislocation did not occur.

CONFLICT OF INTEREST

Declared none.

ACKNOWLEDGEMENT

The authors did not receive and will not receive any benefits or funding from any commercial party related directly or indirectly to the subject of this article.

REFERENCES

[1] Soong M, Rubash HE, Macaulay W. Dislocation after total hip arthroplasty J Am Acad Orthop Surg 2004; 12(5): 314-21.
[2] Su EP, Pellicci PM. The role of constrained liners in total hip arthroplasty Clin Orthop 2004; 420: 122-9.
[3] Morrey BF. Difficult complications after hip joint replacement Dislocation Clin Orthop 1997; 344: 179-87.
[4] Shrader MW, Parvizi J, Lewallen DG. The use of a constrained acetabular component to treat instability after total hip arthroplasty J Bone Joint Surg Am 2003; 85(11): 2179-83.
[5] Berend KR, Lombardi AV Jr, Mallory TH, Adams JB, Russell JH, Groseth KL. The long-term outcome of 755 consecutive constrained acetabular components in total hip arthroplasty examining the successes and failures J Arthroplasty 2005; 20(7 Suppl. 3): 93-102.
[6] Khan RJ, Fick D, Alakeson R, et al. A constrained acetabular component for recurrent dislocation J Bone Joint Surg Br 2006; 88(7): 870-6.
[7] McPherson EJ, Costigan WM, Gerhardt MB, Norris LR. Closed reduction of dislocated total hip with S-ROM constrained acetabular component J Arthroplasty 1999; 14(7): 882-5.
[8] Miller CW, Zura RD. Closed reduction of a dislocation of a constrained acetabular component J Arthroplasty 2001; 16(4): 504-.
[9] Harman MK, Hodge WA, Banks SA. Closed reduction of constrained total hip arthroplasty Clin Orthop 2003; 414: 121-8.
[10] Birdwell S, Wilke E. Closed reduction of constrained total hip arthroplasty in the ED J Emerg Med 2011; 40(2): 162-6.
[11] Robertson WJ, Mattern CJ, Hur J, Su EP, Pellicci PM. Failure mechanisms and closed reduction of a constrained tripolar acetabular liner J Arthroplast 2009; 24(2): 322.e5-11.
[12] Gaines RJ, Hardenbrook M. Closed reduction of a dislocated total hip arthroplasty with a constrained acetabular component Am J Orthop (Belle Mead NJ) 2009; 38(10): 523-.
[13] Flint JH, Phisitkul P, Callaghan JJ. Closed reduction of a dislocated constrained total hip arthroplasty using a novel technique with a peg board Orthopedics 2010; 10: 201-3.

Endorsements



"Open access will revolutionize 21st century knowledge work and accelerate the diffusion of ideas and evidence that support just in time learning and the evolution of thinking in a number of disciplines."


Daniel Pesut
(Indiana University School of Nursing, USA)

"It is important that students and researchers from all over the world can have easy access to relevant, high-standard and timely scientific information. This is exactly what Open Access Journals provide and this is the reason why I support this endeavor."


Jacques Descotes
(Centre Antipoison-Centre de Pharmacovigilance, France)

"Publishing research articles is the key for future scientific progress. Open Access publishing is therefore of utmost importance for wider dissemination of information, and will help serving the best interest of the scientific community."


Patrice Talaga
(UCB S.A., Belgium)

"Open access journals are a novel concept in the medical literature. They offer accessible information to a wide variety of individuals, including physicians, medical students, clinical investigators, and the general public. They are an outstanding source of medical and scientific information."


Jeffrey M. Weinberg
(St. Luke's-Roosevelt Hospital Center, USA)

"Open access journals are extremely useful for graduate students, investigators and all other interested persons to read important scientific articles and subscribe scientific journals. Indeed, the research articles span a wide range of area and of high quality. This is specially a must for researchers belonging to institutions with limited library facility and funding to subscribe scientific journals."


Debomoy K. Lahiri
(Indiana University School of Medicine, USA)

"Open access journals represent a major break-through in publishing. They provide easy access to the latest research on a wide variety of issues. Relevant and timely articles are made available in a fraction of the time taken by more conventional publishers. Articles are of uniformly high quality and written by the world's leading authorities."


Robert Looney
(Naval Postgraduate School, USA)

"Open access journals have transformed the way scientific data is published and disseminated: particularly, whilst ensuring a high quality standard and transparency in the editorial process, they have increased the access to the scientific literature by those researchers that have limited library support or that are working on small budgets."


Richard Reithinger
(Westat, USA)

"Not only do open access journals greatly improve the access to high quality information for scientists in the developing world, it also provides extra exposure for our papers."


J. Ferwerda
(University of Oxford, UK)

"Open Access 'Chemistry' Journals allow the dissemination of knowledge at your finger tips without paying for the scientific content."


Sean L. Kitson
(Almac Sciences, Northern Ireland)

"In principle, all scientific journals should have open access, as should be science itself. Open access journals are very helpful for students, researchers and the general public including people from institutions which do not have library or cannot afford to subscribe scientific journals. The articles are high standard and cover a wide area."


Hubert Wolterbeek
(Delft University of Technology, The Netherlands)

"The widest possible diffusion of information is critical for the advancement of science. In this perspective, open access journals are instrumental in fostering researches and achievements."


Alessandro Laviano
(Sapienza - University of Rome, Italy)

"Open access journals are very useful for all scientists as they can have quick information in the different fields of science."


Philippe Hernigou
(Paris University, France)

"There are many scientists who can not afford the rather expensive subscriptions to scientific journals. Open access journals offer a good alternative for free access to good quality scientific information."


Fidel Toldrá
(Instituto de Agroquimica y Tecnologia de Alimentos, Spain)

"Open access journals have become a fundamental tool for students, researchers, patients and the general public. Many people from institutions which do not have library or cannot afford to subscribe scientific journals benefit of them on a daily basis. The articles are among the best and cover most scientific areas."


M. Bendandi
(University Clinic of Navarre, Spain)

"These journals provide researchers with a platform for rapid, open access scientific communication. The articles are of high quality and broad scope."


Peter Chiba
(University of Vienna, Austria)

"Open access journals are probably one of the most important contributions to promote and diffuse science worldwide."


Jaime Sampaio
(University of Trás-os-Montes e Alto Douro, Portugal)

"Open access journals make up a new and rather revolutionary way to scientific publication. This option opens several quite interesting possibilities to disseminate openly and freely new knowledge and even to facilitate interpersonal communication among scientists."


Eduardo A. Castro
(INIFTA, Argentina)

"Open access journals are freely available online throughout the world, for you to read, download, copy, distribute, and use. The articles published in the open access journals are high quality and cover a wide range of fields."


Kenji Hashimoto
(Chiba University, Japan)

"Open Access journals offer an innovative and efficient way of publication for academics and professionals in a wide range of disciplines. The papers published are of high quality after rigorous peer review and they are Indexed in: major international databases. I read Open Access journals to keep abreast of the recent development in my field of study."


Daniel Shek
(Chinese University of Hong Kong, Hong Kong)

"It is a modern trend for publishers to establish open access journals. Researchers, faculty members, and students will be greatly benefited by the new journals of Bentham Science Publishers Ltd. in this category."


Jih Ru Hwu
(National Central University, Taiwan)


Browse Contents



Webmaster Contact: info@benthamopen.net
Copyright © 2019 Bentham Open