Silicone Gel Bleed on Breast Implants



Felipe Rebello Lourenço*, Irene Satiko Kikuchi , Terezinha de Jesus Andreoli Pinto
Faculty of Pharmaceutical Sciences, University of São Paulo, Brazil

Abstract

Silicone is a polymer widely used for medical purposes, including breast implants. The aim of this work is to verify the occurrence of silicone gel diffusion through the lining membrane (bleed) of gel breast implants with smooth, textured, and polyurethane foam coated surfaces, sterilized by dry heat and ethylene oxide. The results enable us to conclude that there is significant difference for the bleed results among gel breast implants of smooth, textured, and polyurethane foam coated surfaces in relation to the bleed test. The bleed results for implants sterilized by dry heat and ethylene oxide also show significant difference. These results suggest that the gel diffusion through the lining membrane is related to the type of gel breast implants, as well as to the sterilization method. It means that polyurethane foam coated surface breast implant shows less gel diffusion than textured and smooth breast implants and breast implants sterilized by dry heat show less gel diffusion than those sterilied by ethylene oxide

Keywords: Breast implants, silicone, bleed..


Article Information


Identifiers and Pagination:

Year: 2011
Volume: 3
First Page: 14
Last Page: 17
Publisher Id: TOBIOMTJ-3-14
DOI: 10.2174/1876502501103010014

Article History:

Received Date: 16/12/2010
Revision Received Date: 21/1/2011
Acceptance Date: 07/02/2011
Electronic publication date: 18/10/2011
Collection year: 2011

© Lourenço et al.; Licensee Bentham Open.

open-access license: This is an open access article licensed under the terms of the Creative Commons Attribution Non-Commercial License (http: //creativecommons.org/licenses/by-nc/3.0/) which permits unrestricted, non-commercial use, distribution and reproduction in any medium, provided the work is properly cited.


* Address correspondence to this author at the Faculty of Pharmaceutical Sciences, University of São Paulo, Brazil; Tel: 55-11-3091-2218; Fax: 55-11-3091-3626; E-mail: feliperl@usp.br





INTRODUCTION

Silicone is a polymer used in varied medical applications, including breast implants. Intra-capsular leakage to the surrounding tissue and circulation is potentially toxic due to the development of cell and humoral response. The silicone gel implant migration has been appointed as the main cause of capsular contracture [1-6].

The capsular contracture results from a foreign body type of reaction, with the isolation of the implant from the tissue in which it is contained by a fibrous capsule, what can result in breast hardening, thus altering the shape softness and the consistence smoothness, in same cases leading to a new surgery [7-3].

Sterility is a demand in which concerns gel silicone implants and it can be achieved by physical and chemical methods. Among physical sterilization processes, dry heat is rather effective. However, it can lead to silicone gel hydrolysis and consequent damage of the product characteristics, as it is performed at high temperature. Chemical sterilization employing ethylene oxide offers the advantage of effective treatment at relatively low temperatures. However, gas residue, alkylate agents, can react with silicone functional groups, altering its properties [14-19].

Thus the aim of this work is to verify the occurrence of silicone gel diffusion through the lining membrane of gel breast implants of smooth (smooth), textured (text), and polyurethane foam coated (CPU) surfaces, sterilized by dry heat (heat) and ethylene oxide (ETO).

MATERIAL AND METHODS

The bleed test was performed according to ASTM F703-96 norm [20], employing silicone discs with a diameter of 50 mm (surface area of 19,635 cm2), thickness of 3,125mm ± 0,025mm and hardness of 70 ShA, as a support for the implants. 60 gel breast implants with smooth, textured, and coated with polyurethane foam surfaces (Fig. 1), sterilized by dry heat and ethylene oxide, were assayed (Table 1). ASTM F703-96 norm covers the requirements for silicone gel filled, saline inflatable silicone gel filled, and saline inflatable, smooth shell implantable breast prostheses intended for use in surgical reconstruction, augmentation, or replacement of the breast.

Fig. (1)

Gel breast implants with smooth surface (A), textured surface (B) and polyurethane foam coated surface (C).



Table 1

Numbers Gel Breast Implants Assayed for Bleed Test




The silicone discs were previously clean with isopropyl alcohol, being afterwards kept in incubator (Nova Ética, Mod. 411 D, ± 0,1 ºC) for 12 hours, at 43,0 ± 2,0 ºC, for steadiness, being then weighed in analytical scales (Mettler Toledo, AG3000, ± 0,1 mg).

After implant cleaning with isopropyl alcohol and room temperature drying for 2 hours, the implants were placed on the silicone discs (Fig. 2) and put in an oven at 43,0 ± 2,0 ºC for a week, after which the first weighing procedure took place.

Fig. (2)

Gel breast implant and silicone disc.



After this period the control and support discs were removed from the incubator and left at room temperature for steadiness for 1 hour. They were then weighed on analytical scales, being afterwards replaced in the incubator with the implants deposited on the support discs. This procedure was repeated for 8 consecutive weeks. The assay was performed in a room with controlled temperature and relative humidity (26,0 ± 3,0 ºC and 62 ± 16% UR).

RESULTS

The results of the bleed test, performed in an 8-week period, for gel breast implants with smooth, textured, and polyurethane foam coated surfaces, sterilized by dry heat and ethylene oxide, are shown in the graph below (Fig. 3).

Fig. (3)

Results of bleed test for gel breast implants with smooth, textured and polyurethane foam coated surfaces, sterilized by dry heat and ethylene oxide.



The variance analysis (Fig. 4) allows us to conclude that there is significant difference (p = 0,000) between the results obtained for the breast gel implants of smooth, textured, and polyurethane foam coated surfaces in relation to bleed test. Moreover there is also significant difference (p = 0,000) between the results obtained for gel breast implants sterilized by dry heat and ethylene oxide relatively to the bleed test.

Fig. (4)

Analisys of Variance (ANOVA) of bleed test for gel breast implants with smooth, textured and polyurethane foam coated surfaces, sterilized by dry heat and ethylene oxide.



The results indicates that the gel “bleed” is more affect by the type of surface affects than by the sterilization process. Smooth surface breast implants showed a gel “bleed” of 1,0455 mg/cm2/week against 0,3377 mg/cm2/week for textured surface breast implants and 0,1752 mg/cm2/week for polyurethane foam coated surface.

Breast implants sterilized by dry heat showed gel “bleed” of 0,47695 mg/cm2/week against 0,56202 mg/cm2/week for those sterilized by ethylene oxide. Although these values are not so distant, if we evaluate separately each type of surface, the breast implants sterilized by dry heat always showed lower gel “bleed” than those sterilized by ethylene oxide, as we can observe in Fig. (3).

DISCUSSION

The evaluation of the amount of gel diffused by the implants is fundamental, due to the fact that this is the main cause of capsular contracture. Since the first procedures of breast implants several modifications have taken place with a view to the decrease of capsular contracture incidence. However implant evolution has not been paralleled by the development of methodologies to evaluate these materials. Several studies have been carried out to evaluate the safety of implants, mainly involving explants or adjacent tissues to the implants. Nevertheless there is lack of methods to evaluate the rate of gel silicone migration, as the only procedure described under international norm, despite supplying orientation in what concerns comparative studies between different types of implants, does not give specifications as to the acceptable rate of silicone gel migration.

The Guidance for Industry and FDA Staff ─ Saline, Silicone Gel, and Alternative Breast Implants, from 17 November 2006, recommends the study of silicone gel migration employing silicone implants immersed in media which simulate the composition of human interstitial fluids [21]. The medium must be changed, simulating an in vivo dynamic condition and the test must be carried out until gel constituents can be observed in the medium. The Guide also states that the constituents should be identified, as well as the respective rates of migration and the variance of the latter throughout time. The proposed modification is effectively a better simulation of physiological conditions than the method adopted in this work. However its application is not possible due to the lack of information that can, moreover, demonstrate the advantages of its use.

The ASTM F703-96 [20] method employed in this work does not supply information about the nature of the material that migrates from the silicone implants. However the results obtained give important information concerning the amount of gel diffused in relation to the sterilization method and type of coating of the implants. Smooth implants present higher level gel diffusion than textured surface ones, showing the latter a higher level of diffusion than the polyurethane foam coated ones. Also, implants sterilized by dry heat show minor gel diffusion compared with those sterilized by ethylene oxide.

The inflammatory potential risk of silicone breast implants has focused on the gel “bleed” through the membrane [4]. Since the silicone gel and the membrane are manufactured from the same materials, the membrane is not impermeable to its contents [4]. So, breast implants with lower gel “bleed” may show less inflammatory potential.

Smooth breast implants were the firsts silicone gel-filled implants used in aesthetic surgery [22]. This type of breast implant show a higher incidence of post-surgery problems. In the 70’s the firsts breast implants covered with a thin layer of polyurethane foam were developed. Polyurethane-covered implants show a lessened incidence of capsular contracture [22]. One of the reasons for this lower toxicity of covered implants may be explained due to a lower gel “bleed” of this implants. Introduction of barriers layers may improve the smooth surface membrane to lower the gel “bleed” in this kind of breast implant.

Animal studies showed that higher molecular weight silicone are more reactive than lower molecular weight silicone [4]. The method employed in this study do not allow the evaluation of molecular weight of the silicone that “bleed” through the membrane. However, we can concluded that there are significant difference of gel “bleed” among the types of breast implants evaluated.

These results suggest that the diffusion of gel through the coating membrane (bleed) is related to the type of surface gel breast implants have, as well as to the sterilization method employed.

REFERENCES

Track Your Manuscript:


Endorsements



"Open access will revolutionize 21st century knowledge work and accelerate the diffusion of ideas and evidence that support just in time learning and the evolution of thinking in a number of disciplines."


Daniel Pesut
(Indiana University School of Nursing, USA)

"It is important that students and researchers from all over the world can have easy access to relevant, high-standard and timely scientific information. This is exactly what Open Access Journals provide and this is the reason why I support this endeavor."


Jacques Descotes
(Centre Antipoison-Centre de Pharmacovigilance, France)

"Publishing research articles is the key for future scientific progress. Open Access publishing is therefore of utmost importance for wider dissemination of information, and will help serving the best interest of the scientific community."


Patrice Talaga
(UCB S.A., Belgium)

"Open access journals are a novel concept in the medical literature. They offer accessible information to a wide variety of individuals, including physicians, medical students, clinical investigators, and the general public. They are an outstanding source of medical and scientific information."


Jeffrey M. Weinberg
(St. Luke's-Roosevelt Hospital Center, USA)

"Open access journals are extremely useful for graduate students, investigators and all other interested persons to read important scientific articles and subscribe scientific journals. Indeed, the research articles span a wide range of area and of high quality. This is specially a must for researchers belonging to institutions with limited library facility and funding to subscribe scientific journals."


Debomoy K. Lahiri
(Indiana University School of Medicine, USA)

"Open access journals represent a major break-through in publishing. They provide easy access to the latest research on a wide variety of issues. Relevant and timely articles are made available in a fraction of the time taken by more conventional publishers. Articles are of uniformly high quality and written by the world's leading authorities."


Robert Looney
(Naval Postgraduate School, USA)

"Open access journals have transformed the way scientific data is published and disseminated: particularly, whilst ensuring a high quality standard and transparency in the editorial process, they have increased the access to the scientific literature by those researchers that have limited library support or that are working on small budgets."


Richard Reithinger
(Westat, USA)

"Not only do open access journals greatly improve the access to high quality information for scientists in the developing world, it also provides extra exposure for our papers."


J. Ferwerda
(University of Oxford, UK)

"Open Access 'Chemistry' Journals allow the dissemination of knowledge at your finger tips without paying for the scientific content."


Sean L. Kitson
(Almac Sciences, Northern Ireland)

"In principle, all scientific journals should have open access, as should be science itself. Open access journals are very helpful for students, researchers and the general public including people from institutions which do not have library or cannot afford to subscribe scientific journals. The articles are high standard and cover a wide area."


Hubert Wolterbeek
(Delft University of Technology, The Netherlands)

"The widest possible diffusion of information is critical for the advancement of science. In this perspective, open access journals are instrumental in fostering researches and achievements."


Alessandro Laviano
(Sapienza - University of Rome, Italy)

"Open access journals are very useful for all scientists as they can have quick information in the different fields of science."


Philippe Hernigou
(Paris University, France)

"There are many scientists who can not afford the rather expensive subscriptions to scientific journals. Open access journals offer a good alternative for free access to good quality scientific information."


Fidel Toldrá
(Instituto de Agroquimica y Tecnologia de Alimentos, Spain)

"Open access journals have become a fundamental tool for students, researchers, patients and the general public. Many people from institutions which do not have library or cannot afford to subscribe scientific journals benefit of them on a daily basis. The articles are among the best and cover most scientific areas."


M. Bendandi
(University Clinic of Navarre, Spain)

"These journals provide researchers with a platform for rapid, open access scientific communication. The articles are of high quality and broad scope."


Peter Chiba
(University of Vienna, Austria)

"Open access journals are probably one of the most important contributions to promote and diffuse science worldwide."


Jaime Sampaio
(University of Trás-os-Montes e Alto Douro, Portugal)

"Open access journals make up a new and rather revolutionary way to scientific publication. This option opens several quite interesting possibilities to disseminate openly and freely new knowledge and even to facilitate interpersonal communication among scientists."


Eduardo A. Castro
(INIFTA, Argentina)

"Open access journals are freely available online throughout the world, for you to read, download, copy, distribute, and use. The articles published in the open access journals are high quality and cover a wide range of fields."


Kenji Hashimoto
(Chiba University, Japan)

"Open Access journals offer an innovative and efficient way of publication for academics and professionals in a wide range of disciplines. The papers published are of high quality after rigorous peer review and they are Indexed in: major international databases. I read Open Access journals to keep abreast of the recent development in my field of study."


Daniel Shek
(Chinese University of Hong Kong, Hong Kong)

"It is a modern trend for publishers to establish open access journals. Researchers, faculty members, and students will be greatly benefited by the new journals of Bentham Science Publishers Ltd. in this category."


Jih Ru Hwu
(National Central University, Taiwan)


Browse Contents




Webmaster Contact: info@benthamopen.net
Copyright © 2023 Bentham Open